SURGICAL STAINLESS STEEL SUTURE
Report
- Report Number
- 2210968-2013-02807
- Event Type
- Injury
- Date Received
- March 26, 2013
- Date of Event
- March 1, 2013
- Report Date
- March 3, 2013
- Manufacturer
- ETHICON, INC.
- Product Code
- GAQ
- PMA / PMN Number
- K946173
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). CONCLUSION: THE ACTUAL DEVICE INVOLVED IN THIS EVENT WAS RETURNED FOR EVALUATION. A VISUAL INSPECTION OF THE SCANNING ELECTRON MICROSCOPE (SEM) PHOTOS WAS PERFORMED WHICH REVEALED INDENTS AND SCUFF MARKS AROUND THE BREAK AREA PRODUCED DURING HANDLING BY A SURGICAL NEEDLE HOLDERS OR SOME OTHER GRIPPING DEVISE. THE NEEDLES FRACTURED DUE TO TENSILE OVERLOAD GENERATED DURING SEVERE MECHANICAL DEFORMATION, WITH SIGNS OF DUCTILE FAILURE. THERE ARE NO SIGNS OF MANUFACTURING DEFECTS FOUND ON THE NEEDLES. IN ORDER TO AVOID THIS KIND OF DAMAGE THE PACKAGE INSERT CAUTIONS: TO AVOID DAMAGING NEEDLE POINTS AND SWAGE AREAS, GRASP THE NEEDLE IN AN AREA ONE-THIRD (1/3) TO ONE-HALF (1/2) OF THE DISTANCE FROM THE SWAGED END TO THE POINT.
(B)(4): THE PRODUCT UPON WHICH THIS MEDWATCH IS BASED IS ANTICIPATED. ONCE THE PRODUCT IS RECEIVED, ANY FURTHER INFORMATION DERIVED FROM THE EVALUATION WILL BE SUBMITTED IN A SUPPLEMENTAL 3500A FORM.IN ADDITION, A REVIEW OF THE BATCH MANUFACTURING RECORDS WAS CONDUCTED AND THE BATCH MET ALL FINISHED GOODS RELEASE CRITERIA. THIS IS ONE OF TWO MEDWATCHES BEING SUBMITTED AS TWO DEVICES WERE INVOLVED IN THIS EVENT. SEE MEDWATCH 2210968-2013-02806. THE SAME PATIENT IS REPRESENTED IN EACH MEDWATCH.
IT WAS REPORTED THAT A PATIENT UNDERWENT A SURGICAL PROCEDURE FOR CONSTRICTIVE PERICARDITIS ON (B)(6) 2013 AND SUTURE WAS USED. WHEN THE SURGEON WAS STITCHING THE STERNUM, THE NEEDLE TIP BROKE. THE SURGEON COULD NOT FIND THE NEEDLEPOINT. CHANGED THE SECOND ONE TO CONTINUE THE PROCEDURE. IT IS UNKNOWN HOW THE PROCEDURE WAS COMPLETED. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 124758 | SURGICAL STAINLESS STEEL SUTURE | SUTURE, NON-ABSORBABLE | GAQ | ETHICON, INC. | NA | EDE203 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 57 YR | Other |