FDA Adverse Event Malfunction Summary report: N

CORE IMPACTION DRILL

MDR report key: 3021591 · Received March 26, 2013

Report

Report Number
0001811755-2013-00604
Event Type
Malfunction
Date Received
March 26, 2013
Date of Event
February 28, 2013
Report Date
February 28, 2013
Manufacturer
STRYKER INSTRUMENTS-KALAMAZOO
Product Code
DZJ
PMA / PMN Number
K040369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

UPON DISASSEMBLY FOR VISUAL INSPECTION CORROSION WAS FOUND ON THE BEARING, NOSE CONE AND SPINDLE HOUSING. ADDITIONALLY THE ROTOR AND THE MOTOR WERE SEIZED TOGETHER.

Description of Event or Problem · 1

WHILE TESTING THE CORE IMPACTION DRILL, IT WAS REPORTED THAT THE DEVICE HEATED UP. THIS EVENT DID NOT OCCUR DURING A SURGICAL PROCEDURE, SO THERE WAS NO PATIENT INVOLVEMENT. THERE WAS NO USER INJURY REPORTED AND NO ADVERSE CONSEQUENCES ALLEGED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
123627 CORE IMPACTION DRILL DRIVER, WIRE, AND BONE DRILL, MANUAL DZJ STRYKER INSTRUMENTS-KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1