FDA Adverse Event
Malfunction
Summary report: N
CORE IMPACTION DRILL
MDR report key: 3021591
·
Received March 26, 2013
Report
- Report Number
- 0001811755-2013-00604
- Event Type
- Malfunction
- Date Received
- March 26, 2013
- Date of Event
- February 28, 2013
- Report Date
- February 28, 2013
- Manufacturer
- STRYKER INSTRUMENTS-KALAMAZOO
- Product Code
- DZJ
- PMA / PMN Number
- K040369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
UPON DISASSEMBLY FOR VISUAL INSPECTION CORROSION WAS FOUND ON THE BEARING, NOSE CONE AND SPINDLE HOUSING. ADDITIONALLY THE ROTOR AND THE MOTOR WERE SEIZED TOGETHER.
Description of Event or Problem · 1
WHILE TESTING THE CORE IMPACTION DRILL, IT WAS REPORTED THAT THE DEVICE HEATED UP. THIS EVENT DID NOT OCCUR DURING A SURGICAL PROCEDURE, SO THERE WAS NO PATIENT INVOLVEMENT. THERE WAS NO USER INJURY REPORTED AND NO ADVERSE CONSEQUENCES ALLEGED WITH THIS EVENT.