FDA Adverse Event Malfunction Summary report: N

TAXUS¿ ELEMENT¿

MDR report key: 3021533 · Received March 26, 2013

Report

Report Number
2134265-2013-01628
Event Type
Malfunction
Date Received
March 26, 2013
Date of Event
November 7, 2012
Report Date
February 25, 2013
Manufacturer
BOSTON SCIENTIFIC - GALWAY
Product Code
NIQ
PMA / PMN Number
P100023
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

AGE AT TIME OF EVENT: 18 YEARS OR OLDER. (B)(4). DEVICE IS A COMBINATION PRODUCT. DEVICE EVALUATED BY MANUFACTURER - INITIAL VISUAL EXAMINATION OF THE RETURNED DEVICE, FOUND PROXIMAL STENT STRUT DAMAGE PRESENT, AS A NUMBER OF STRUTS WERE BENT BACK DISTALLY. THE BALLOON AND TIP SECTIONS OF THE DEVICE WERE VISUALLY AND MICROSCOPICALLY EXAMINED AND NO ISSUES WERE NOTED WITH THEIR PROFILES THAT COULD HAVE POTENTIALLY CONTRIBUTED TO THE COMPLAINT INCIDENT. THE BALLOON WAS TIGHTLY WRAPPED AND WAS NOT SUBJECTED TO POSITIVE PRESSURE. NO FURTHER DAMAGE WAS NOTED WITH THE RETURNED DEVICE WHICH COULD HAVE CONTRIBUTED TO THE REPORTED COMPLAINT. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL/PROCEDURAL FACTORS. (B)(4).

Description of Event or Problem · 1

THIS COMPLAINT IS NOW REPORTABLE BASED ON THE ANALYSIS COMPLETED ON (B)(6) 2013. IT WAS REPORTED THAT DURING A STENTING TREATMENT PROCEDURE, THE 2.50X24MM TAXUS ELEMENT STENT WOULD NOT CROSS THE LESION. ACCESS WAS OBTAINED VIA THE FEMORAL ARTERY. THE 90% STENOSED, 24MM IN LENGTH, LESION BEING TREATED WAS LOCATED IN THE SEVERELY TORTUOUS AND SEVERELY CALCIFIED LEFT ANTERIOR DESCENDING (LAD) ARTERY. THE LESION WAS PREDILATED WITH AN UNKNOWN SEMI COMPLIANT BALLOON. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS IS STABLE. HOWEVER, THE PRODUCT ANALYSIS REVEALED STENT DAMAGE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
123254 TAXUS¿ ELEMENT¿ CORONARY DRUG-ELUTING STENT NIQ BOSTON SCIENTIFIC - GALWAY H7493902524250 15027015

Patients

Seq Age Sex Outcome Treatment
1