FDA Adverse Event Malfunction Summary report: N

MICRO SAGITTAL SAW

MDR report key: 3021379 · Received March 26, 2013

Report

Report Number
0001811755-2013-00603
Event Type
Malfunction
Date Received
March 26, 2013
Date of Event
February 27, 2013
Report Date
February 27, 2013
Manufacturer
STRYKER INSTRUMENTS-KALAMAZOO
Product Code
ERL
PMA / PMN Number
K040300
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

UPON DISASSEMBLY FOR VISUAL EXAMINATION, WIDESPREAD CORROSION WAS DISCOVERED.

Description of Event or Problem · 1

THE MICRO SAGGITAL SAW WAS RETURNED FOR SERVICE. UPON EVALUATION AT THE MANUFACTURER, IT DISPLAYED A BIAS CURRENT ERROR WHEN CONNECTED TO THE CONSOLE SIGNALING A CONDTIION OCCURRED FROM WHICH THE DEVICE HAS THE POTENTIAL TO RUN WITHOUT USER ACTIVATION. THERE WAS NO PATIENT INVOLVEMENT, AND THERE WERE NO USER INJURIES OR ADVERSE CONSEQUENCES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
122937 MICRO SAGITTAL SAW DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE ERL STRYKER INSTRUMENTS-KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1