FDA Adverse Event Injury Summary report: N

PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM

MDR report key: 3021150 · Received March 26, 2013

Report

Report Number
2024168-2013-01715
Event Type
Injury
Date Received
March 26, 2013
Date of Event
March 7, 2013
Report Date
March 7, 2013
Manufacturer
AV-TEMECULA-CT
Product Code
MGB
PMA / PMN Number
P960043
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. THE DEVICE WAS RECEIVED. INVESTIGATION IS NOT YET COMPLETE. A FOLLOW UP REPORT WILL BE SUBMITTED WITH ALL ADDITIONAL RELEVANT INFORMATION. THE SECOND PROGLIDE DEVICE REFERENCED IS BEING FILED UNDER A SEPARATE MEDWATCH MFR NUMBER.

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED FOR EVALUATION. THE REPORTED EVENT WAS NOT CONFIRMED. THE LINK HAD DETACHED AT THE POSTERIOR CUFF. A LINK DETACHMENT HAS THE SAME RESULT AND APPEARANCE AS THE REPORTED NEEDLE TO CUFF MISS. BASED ON VISUAL AND FUNCTIONAL ANALYSIS OF THE RETURNED DEVICE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THE REPORTED EVENT. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY.

Description of Event or Problem · 1

IT WAS REPORTED THAT AN ARTERIOTOMY CLOSURE OF THE RIGHT COMMON FEMORAL ARTERY WAS ATTEMPTED USING A PROGLIDE DEVICE AFTER A CARDIAC CATHETERIZATION INTERVENTIONAL PROCEDURE. REPORTEDLY, A CUFF MISS OCCURRED. A SECOND PROGLIDE DEVICE WAS USED WITH THE SAME RESULTS. A THIRD PROGLIDE DEVICE WAS USED TO ACHIEVE HEMOSTASIS. THERE WAS NO REPORTED ADVERSE PATIENT SEQUELA OR CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE OR THERAPY. THE PHYSICIAN IS REPORTEDLY TRAINED IN THE USE OF THE PROGLIDE DEVICE. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
124520 PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM SUTURE MEDIATED CLOSURE MGB AV-TEMECULA-CT 30108J1

Patients

Seq Age Sex Outcome Treatment
1 67 YR Required Intervention SHEATH: 6F ANGIOMAX