FDA Adverse Event Malfunction Summary report: N

HS III PROXIMAL SEAL SYSTEM 4.3MM

MDR report key: 3021008 · Received March 20, 2013

Report

Report Number
2242352-2013-00266
Event Type
Malfunction
Date Received
March 20, 2013
Date of Event
March 1, 2013
Report Date
March 1, 2013
Manufacturer
MAQUET CARDIOVASCULAR, LLC
Product Code
DXC
PMA / PMN Number
K080169
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WV, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION RESULTS: DEVICE 1: THE DEVICE WAS RETURNED TO THE FACTORY FOR EVALUATION. IT SHOWED SIGNS OF CLINICAL USAGE. THERE WAS SOME EVIDENCE OF BLOOD ON THE DELIVERY DEVICE, INSIDE OF THE DELIVERY TUBE AND ON THE SEAL. THE LOADING DEVICE WAS NOT RETURNED. THE TENSION SPRING ASSEMBLY WAS LOCATED INSIDE OF THE DELIVERY DEVICE TUBE; IT REMAINED ANCHORED TO THE TENSION SPRING ASSEMBLY. THE GREEN SLIDE LOCK WAS UNLOCKED AND THE WHITE PLUNGER WAS DEPRESSED ON THE DELIVERY DEVICE. BASED UPON THE EVALUATION RESULTS, THE REPORTED COMPLAINT COULD NOT BE CONFIRMED FOR FAILURE TO LOAD; HOWEVER, IT WAS CONFIRMED FOR FAILURE TO DEPLOY. DEVICE 2: THE DEVICE WAS RETURNED TO THE FACTORY FOR EVALUATION. IT SHOWED SIGNS OF CLINICAL USAGE. THERE WAS A SIGNIFICANT AMOUNT OF BLOOD ON THE DELIVERY DEVICE, INSIDE OF THE DELIVERY TUBE AND ON THE SEAL. THE LOADING DEVICE WAS NOT RETURNED. THE TENSION SPRING ASSEMBLY WAS LOCATED INSIDE OF THE DELIVERY DEVICE TUBE; IT REMAINED ANCHORED TO THE TENSION SPRING ASSEMBLY. THE GREEN SLIDE LOCK WAS UNLOCKED AND THE WHITE PLUNGER WAS DEPRESSED ON THE DELIVERY DEVICE. BASED UPON THE EVALUATION RESULTS, THE REPORTED COMPLAINT COULD NOT BE CONFIRMED FOR FAILURE TO LOAD; HOWEVER, IT WAS CONFIRMED FOR FAILURE TO DEPLOY. (B)(4).

Description of Event or Problem · 1

THE HOSP REPORTED THAT DURING A CORONARY ARTERY BYPASS PROCEDURE, TWO HEARTSTRING III SEAL FAILED TO LOAD PROPERLY. A REPLACEMENT DEVICE WAS USED TO COMPLETE THE PROCEDURE. THE HOSP DID NOT REPORT ANY PT EFFECTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
116947 HS III PROXIMAL SEAL SYSTEM 4.3MM CLAMPLESS BEATING HEART DXC MAQUET CARDIOVASCULAR, LLC HSK-3043 ASKU

Patients

Seq Age Sex Outcome Treatment
1 NA