FDA Adverse Event Death Summary report: N

COMPONENT MONITORING SYSTEM

MDR report key: 302092 · Received October 26, 2000

Report

Report Number
9610816-2000-00207
Event Type
Death
Date Received
October 26, 2000
Date of Event
October 26, 2000
Manufacturer
AGILENT TECHNOLOGIES DEUTSCHLAND GMBH
Product Code
DPS
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
LA, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE ATTENDING NURSE SUSPENDED THE ALARMS DURING PT CARE. THE NURSE WAS UNAWARE THAT THE ALARMS WERE CONFIGURED FOR INFINITE ALARM SUSPEND. THE PT'S RHYTHM CHANGED TO ASYSTOLE AND NO ALARM WAS GENERATED AS A RESULT OF THE BEDSIDE ALARM BEING SUSPENDED INDEFINITELY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COMPONENT MONITORING SYSTEM CMS DPS AGILENT TECHNOLOGIES DEUTSCHLAND GMBH M1176A NA

Patients

Seq Age Sex Outcome Treatment
1 UNKNOWN Death