FDA Adverse Event
Malfunction
Summary report: N
9800
MDR report key: 3020899
·
Received March 18, 2013
Report
- Report Number
- 1720753-2013-03887
- Event Type
- Malfunction
- Date Received
- March 18, 2013
- Date of Event
- March 13, 2013
- Report Date
- March 18, 2013
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE REPORTED ISSUE COULD NOT BE DUPLICATED. THE BOARDS AND CONNECTORS WERE RESEATED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THE SYSTEM MONITOR POPPED, AND THEN SMELLED A BURNING SMELL FROM THE WORKSTATION. THE SYSTEM WENT DOWN AND HAD TO REBOOT DURING A PROCEDURE. THERE IS NO REPORT OF PT INJURY OR DEATH.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 111810 | 9800 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9800 |