FDA Adverse Event Malfunction Summary report: N

840 VENTILATOR

MDR report key: 3020845 · Received March 18, 2013

Report

Report Number
8020893-2013-00647
Event Type
Malfunction
Date Received
March 18, 2013
Date of Event
February 20, 2013
Report Date
February 20, 2013
Manufacturer
COVIDIEN, FORMERLY NELLCOR PURITAN BENNETT
Product Code
CBK
PMA / PMN Number
K970460
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
NOT APPLICABLE

Narratives

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT AN 840 VENTILATOR HAD A BLANK SCREEN. NO PT INVOLVEMENT. THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) VERIFIED THE MALFUNCTION. THE CSE INSPECTED THE DEVICE AND REPLACED THE GUI BACKLIGHT PCB AND GUI LCD PANEL. THE UNIT PASSED EXTENDED SELF-TESTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
112779 840 VENTILATOR VENTILATOR CBK COVIDIEN, FORMERLY NELLCOR PURITAN BENNETT 840

Patients

Seq Age Sex Outcome Treatment
1