FDA Adverse Event
Malfunction
Summary report: N
840 VENTILATOR
MDR report key: 3020845
·
Received March 18, 2013
Report
- Report Number
- 8020893-2013-00647
- Event Type
- Malfunction
- Date Received
- March 18, 2013
- Date of Event
- February 20, 2013
- Report Date
- February 20, 2013
- Manufacturer
- COVIDIEN, FORMERLY NELLCOR PURITAN BENNETT
- Product Code
- CBK
- PMA / PMN Number
- K970460
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT AN 840 VENTILATOR HAD A BLANK SCREEN. NO PT INVOLVEMENT. THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) VERIFIED THE MALFUNCTION. THE CSE INSPECTED THE DEVICE AND REPLACED THE GUI BACKLIGHT PCB AND GUI LCD PANEL. THE UNIT PASSED EXTENDED SELF-TESTING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 112779 | 840 VENTILATOR | VENTILATOR | CBK | COVIDIEN, FORMERLY NELLCOR PURITAN BENNETT | 840 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |