FDA Adverse Event Malfunction Summary report: N

ACCU-CHEK ® SPIRIT

MDR report key: 3020683 · Received March 26, 2013

Report

Report Number
2183996-2013-00509
Event Type
Malfunction
Date Received
March 26, 2013
Date of Event
March 13, 2013
Report Date
May 22, 2013
Manufacturer
ROCHE DIABETES CARE AG
Product Code
LZG
PMA / PMN Number
K060876
Removal / Correction Number
Z-1646-2009
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

IT WAS UNKNOWN IF THE INITIAL REPORTER SENT REPORT TO THE FDA.

Additional Manufacturer Narrative · 1

THE SOFT COMPONENTS OF THE HOUSING ARE WORN DOWN HEAVILY AND RESTRICTED HANDLING OF THE PUMP IS A POSSIBLE IMPLICATION. THEREFORE, MOISTURE ENTERED THE INSULIN PUMP AND DESTROYS THE FUNCTIONALITY OF THE BUTTONS AND THE PUMP ELECTRONICS. THE UP BUTTON DOES NOT RESPOND TO COMMANDS AND ALSO GIVE NO AUDIBLE FEEDBACK DUE TO THE CONTAMINATION INSIDE THE BUTTON. THE PROBLEM MENTIONED BY THE CUSTOMER IS RELATED TO CARELESS HANDLING OF THE PRODUCT.

Description of Event or Problem · 1

ON (B)(6) 2013, THE PATIENT'S MOTHER REPORTED THAT THE UP BUTTON ON HIS INFUSION DEVICE HAD NOT BEEN FUNCTIONING SINCE (B)(6) 2013. NO PHYSIOLOGICAL EFFECTS WERE REPORTED. THE PATIENT DID NOT REQUIRE MEDICAL ASSISTANCE FROM A HEALTHCARE PROFESSIONAL OR SECOND PARTY TO ADDRESS THE ISSUE. INFUSION DEVICE WAS REPLACED AND WAS REQUESTED TO BE RETURNED FOR PRODUCT EVALUATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
124036 ACCU-CHEK ® SPIRIT INSULIN INFUSION PUMP LZG ROCHE DIABETES CARE AG NA NA

Patients

Seq Age Sex Outcome Treatment
1 020 YR NOVALOG