FDA Adverse Event Malfunction Summary report: N

EMERGE¿

MDR report key: 3020646 · Received March 26, 2013

Report

Report Number
2134265-2013-01692
Event Type
Malfunction
Date Received
March 26, 2013
Date of Event
February 25, 2013
Report Date
February 25, 2013
Manufacturer
BOSTON SCIENTIFIC - MAPLE GROVE
Product Code
LOX
PMA / PMN Number
K113220
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

AGE AT TIME OF EVENT: 18 YEARS OR OLDER. (B)(4). DEVICE EVALUATED BY MANUFACTURER: THE DEVICE HAS NOT BEEN RECEIVED FOR ANALYSIS. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL / PROCEDURAL FACTORS. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PERCUTANEOUS CORONARY INTERVENTION PROCEDURE, A BALLOON RUPTURE OCCURRED. THE PROCEDURE TREATED THE 90% STENOSED TARGET LESION LOCATED IN THE MODERATELY CALCIFIED AND MODERATELY TORTUOUS DISTAL RIGHT CORONARY ARTERY. A NON-BSC GUIDE WIRE WAS ADVANCED AND CROSSED THE LESION. NEXT A 2.5X15MM EMERGE BALLOON CATHETER WAS USED TO DILATE THE LESION, HOWEVER, THE BALLOON RUPTURED AT NOMINAL PRESSURE. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT STATUS IS GOOD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
124798 EMERGE¿ CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS LOX BOSTON SCIENTIFIC - MAPLE GROVE H7493919315250 15661577

Patients

Seq Age Sex Outcome Treatment
1 SION (ASAHI INTECC) GUIDE WIRE| 7FR MACH 1 GUIDE CATHETER