LIBERTY CYCLER
Report
- Report Number
- 2937457-2013-00030
- Event Type
- Other
- Date Received
- March 19, 2013
- Date of Event
- February 17, 2013
- Report Date
- February 18, 2013
- Manufacturer
- FRESENIUS MEDICAL CARE NORTH AMERICA
- Product Code
- FKX
- PMA / PMN Number
- K043363
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
A MEDICAL REVIEW WAS PERFORMED ON THE INFO PROVIDED. A LARGE INTRA-PERITONEAL VOLUME DRAINED AT DRAIN 1 WITH AN UNDETERMINED CAUSE. THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS REPORTED LARGE DRAIN VOLUME. THE DEVICE IS CURRENTLY UNDERGOING EVALUATION AND A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
PT CALLED TECH SUPPORT REGARDING FILLING SLOW DURING A CCPD TREATMENT. PT REPORTED FEELING FULL AFTER THE FILL. PT PERFORMED A STAT DRAIN AND OBTAINED 4745 ML. TREATMENT PRESCRIPTION: 4 FILLS OF 2305 MIL AND NO LAST FILL. NO MID-DAY MANUAL EXCHANGES. TREATMENT DATA OBTAINED FROM THE PT: DRAIN 0: 188 ML. FILL 1: 2296 ML. NINE MINUTES INTO DWELL PATIENT DID A STAT DRAIN DUE TO FEELING FULL, BLOATED, SWOLLEN, AND HAD DIFFICULTY BREATHING. DRAIN 1: 4745 ML. PT CANCELLED THE TREATMENT. PT REPORTED FEELING BETTER AFTER DRAINING AND HAD NO ADVERSE EFFECTS AS A RESULT. PATIENT USES TWO 5 LITER PD SOLUTION BAGS FOR THE TREATMENT. PATIENT REPORTS THE HEATER BAG WAS EMPTY AND THE SUPPLY BAG HAD A LITTLE BIT OF FLUID. OF THE 4 DRAIN BAGS USED, ONE WAS EMPTY AND THE OTHER THREE WERE LESS THAN HALF FULL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 113418 | LIBERTY CYCLER | PERITONEAL DIALYSIS CYCLER | FKX | FRESENIUS MEDICAL CARE NORTH AMERICA | LIBERTY CYCLER |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | PD CATHETER WITH EXTENSION SET| LIBERTY CYCLER TUBING SET| PD SOLUTION |