OT VERIO IQ METER
Report
- Report Number
- 3008382007-2013-06176
- Event Type
- Malfunction
- Date Received
- March 26, 2013
- Report Date
- February 26, 2013
- Manufacturer
- LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
- Product Code
- NBW
- PMA / PMN Number
- K110637
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
FOLLOW-UP #1 (6/19/2013) -DEVICE EVALUATION: THE METER INVOLVED WITH THIS COMPLAINT HAS BEEN RETURNED AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS ON 6/11/2013 WITH THE FOLLOWING FINDINGS: THE METER FAILED TESTING AND WAS FOUND TO HAVE AN OPEN/SHORTED CAPACITOR. IN ADDITION, THE AC ADAPTER WAS TESTED AND PASSED TESTING; THE COMPLAINT WAS NOT CONFIRMED. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.
THE LIFESCAN PRODUCT HAS BEEN RETURNED TO LIFESCAN. EVALUATION HAS NOT YET BEEN COMPLETED. AFTER THE EVALUATION IS COMPLETED, A SUPPLEMENTAL REPORT WILL BE FILLED. NO CONCLUSIONS CAN BE DRAWN AT THIS TIME.
ON (B)(6) 2013, THE REPORTER CONTACTED LIFESCAN USA, ALLEGING BATTERY INDICATOR. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 123604 | OT VERIO IQ METER | GLUCOSE MONITORING SYS/KIT | NBW | LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |