FDA Adverse Event Malfunction Summary report: N

OT VERIO IQ METER

MDR report key: 3020377 · Received March 26, 2013

Report

Report Number
3008382007-2013-06148
Event Type
Malfunction
Date Received
March 26, 2013
Report Date
February 25, 2013
Manufacturer
LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
Product Code
NBW
PMA / PMN Number
K110637
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE LIFESCAN PRODUCT HAS BEEN RETURNED TO LIFESCAN. EVALUATION HAS NOT YET BEEN COMPLETED. AFTER THE EVALUATION IS COMPLETED, A SUPPLEMENTAL REPORT WILL BE FILLED. NO CONCLUSIONS CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

ON (B)(6) 2013, THE REPORTER CONTACTED LIFESCAN (B)(4), ALLEGING BATTERY INDICATOR. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
124177 OT VERIO IQ METER GLUCOSE MONITORING SYS/KIT NBW LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL 3406114

Patients

Seq Age Sex Outcome Treatment
1 80 YR