FDA Adverse Event Injury Summary report: N

TALENT THORACIC STENT GRAFT SYSTEM

MDR report key: 3020085 · Received March 25, 2013

Report

Report Number
2953200-2013-00536
Event Type
Injury
Date Received
March 25, 2013
Date of Event
April 12, 2013
Report Date
June 3, 2013
Manufacturer
MEDTRONIC IRELAND
Product Code
MIH
PMA / PMN Number
P070007
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

METHOD: (FILM). RESULTS: PATIENT¿S CONDITION AFFECTED EFFECTIVENESS OF DEVICE (TYPE II ENDOLEAK). CONCLUSION: DEVICE FAILURE/LACK OF EFFECTIVENESS RELATED TO PATIENT CONDITION (TYPE II ENDOLEAK).

Additional Manufacturer Narrative · 1

(B)(4). RESULTS: ENDOLEAK. DISEASE PROGRESSION. CONCLUSION: DISEASE PROGRESSION.

Description of Event or Problem · 1

A TALENT STENT GRAFT SYSTEM WAS IMPLANTED IN A PATIENT FOR THE ENDOVASCULAR TREATMENT OF 6.5 CM DIAMETER THORACIC AORTIC ANEURYSM IN ZONE 4 APPROXIMATELY 26 MONTHS AGO. THE PROXIMAL NECK WAS 32-30 MM IN DIAMETER. THE COMMON ILIAC ARTERIES WERE MODERATELY CALCIFICATION. THE PROXIMAL AND DISTAL AORTIC NECK HAD MILD THROMBUS. AT THE TWO YEAR FOLLOW-UP A PROXIMAL TYPE I ENDOLEAK WAS SEEN ON THE CT SCAN. THE ANEURYSM EXPANDED 9 MM IN THE PAST YEAR. IT WAS REPORTED THAT THE CAUSE OF THE EVENT WAS MOST LIKELY DUE TO EXPANSION OF THE AORTA. THE PATIENT WILL BE MONITORED. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED AND THE PATIENT IS FINE.

Description of Event or Problem · 1

ADDITIONAL INFORMATION WAS RECEIVED AS FOLLOWS: IT WAS REPORTED THAT THERE WAS ANEURYSM ENLARGEMENT DUE TO AN ENDOLEAK AND A DEVELOPMENT OF A NEW ANEURYSM PROXIMAL TO THE TALENT STENT GRAFT. THE PROXIMAL PORTION OF THE THORACIC AORTA WAS CONVERTED WITH AN ARTIFICIAL GRAFT FOLLOWED BY IMPLANT OF TWO VALIANT STENT GRAFTS WHICH WERE IMPLANTED TO COVER THE TWO PREVIOUSLY IMPLANTED TALENT. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED AND THE PATIENT IS STABLE.

Description of Event or Problem · 1

FILMS FROM 2 YEARS POST-IMPLANT WERE RECEIVED AND REVIEWED AS FOLLOWS: SIGNIFICANT CALCIFICATION WAS OBSERVED WITHIN THE AORTIC ARCH. THE STENT GRAFT WAS POSITIONED IN THE DESCENDING THORACIC; JUST DISTAL TO THE ARCH. CONTRAST WAS SEEN NEAR THE MID-LENGTH OF THE STENT GRAFTS (OVERLAPPING SECTION). THIS MAY BE A TYPE III ENDOLEAK; COULD NOT DETERMINE IF A JUNCTION OR FABRIC TYPE III ENDOLEAK. THERE IS ALSO A POSSIBLE TYPE II ENDOLEAK. NO PROXIMAL OR DISTAL TYPE I COULD BE SEEN CONFIRMED THE PROXIMAL STENT GRAFT OD WAS 37X38MM, AND THE DISTAL OD WAS 30X33MM. THE CAUSE OF THE TYPE III ENDOLEAK COULD NOT BE DETERMINED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
122602 TALENT THORACIC STENT GRAFT SYSTEM SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT MIH MEDTRONIC IRELAND

Patients

Seq Age Sex Outcome Treatment
1 00079 YR Required Intervention