FDA Adverse Event Malfunction Summary report: N

MINICAP TRANSFER SET

MDR report key: 3020017 · Received March 25, 2013

Report

Report Number
1416980-2013-07148
Event Type
Malfunction
Date Received
March 25, 2013
Date of Event
February 20, 2013
Report Date
March 4, 2013
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
KDJ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NO
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). A EVALUATION OF THE ACTUAL SAMPLE WAS CONDUCTED AND ISSUES WERE FOUND RELATED TO THE REPORTED CONDITION DURING THE EVALUATION. VISUAL INSPECTION PERFORMED WITH THE FOLLOWING ISSUES NOTED: BROKEN OCCLUDER FEET. LEAK TESTING PERFORMED WITH NO ISSUES NOTED. CLEAR PASSAGE TEST PERFORMED WITH NO ISSUES NOTED. CLAMP FUNCTION TEST PERFORMED WITH THE FOLLOWING ISSUES NOTED: BROKEN OCCLUDER FEET. INITIAL EVALUATION CONFIRMED THE REPORTED CONDITION. UPON COMPLETION OF BAXTER'S INVESTIGATION, A FOLLOW-UP WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). (B)(6). A BATCH REVIEW WAS CONDUCTED AND NO ISSUES WERE FOUND RELATED TO THE REPORTED CONDITION DURING THE MANUFACTURE OF THE LOT. THE SAMPLE HAS BEEN RECEIVED BY BAXTER, AND THE EVALUATION HAS NOT YET BEEN COMPLETED. A FOLLOW-UP MDR WILL BE SUBMITTED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL RELEVANT INFORMATION IS RECEIVED.

Additional Manufacturer Narrative · 1

(B)(4). IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED A FOLLOW-UP WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED TO BAXTER (B)(4) THAT THERE WAS A LEAK THAT OCCURED AT THE TOP OF THE TRANSFER SET ALTHOUGH THE CLAMP WAS CLOSED. THE PATIENT PUT ANOTHER CLAMP ON THE SET AND PUT ON THE MINICAP. THEN THE PATIENT CONTACTED THE HOSPITAL FOR AN EXCHANGE OF THE SET. THE SET WAS INITIALLY COUPLED TO THE PATIENT TWO MONTHS PRIOR. THERE WAS PATIENT INVOLVEMENT BUT NO REPORTED INJURY OR MEDICAL INTERVENTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
122866 MINICAP TRANSFER SET SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE KDJ BAXTER HEALTHCARE - MOUNTAIN HOME H12B20018

Patients

Seq Age Sex Outcome Treatment
1 HOME CHOICE