FDA Adverse Event Malfunction Summary report: N

CEP 8 (D8Z2) SPECTRUMGREEN PROBE

MDR report key: 3019754 · Received March 25, 2013

Report

Report Number
3005248192-2013-00009
Event Type
Malfunction
Date Received
March 25, 2013
Date of Event
March 4, 2013
Report Date
March 4, 2013
Manufacturer
ABBOTT MOLECULAR, INC.
Product Code
MVU
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

AN ABBOTT MOLECULAR INVESTIGATION CONFIRMED THIS ISSUE AS A PRODUCT MALFUNCTION. (B)(4) WAS OPENED AS A RESULT OF THIS MALFUNCTION ISSUE AND IS IN PROGRESS. ABBOTT MOLECULAR IS PREPARING AN 806 FIELD ACTION TO NOTIFY AFFECTED CUSTOMERS. PREVIOUSLY REPORTED MDRS RELATED TO THIS SAME MALFUNCTION ISSUE: 3005248192-2013-00005, 3005248192-2013-00006, 3005248192-2013-00007, 3005248192-2013-00008.

Description of Event or Problem · 1

THE CEP 8 SPECTRUMGREEN (SG) ASR PROBE KIT, 20UL (LIST 06J37-018) IS AN ANALYTE SPECIFIC REAGENT (ASR) CONTAINING A SINGLE VIAL OF CEP 8 SPECTRUM GREEN PROBE AND A CERTIFICATE OF ANALYSIS (COA). CHROMOSOME ENUMERATION PROBES (CEP) ARE CHROMOSOME-SPECIFIC FISH PROBES THAT HYBRIDIZE TO HIGHLY REPETITIVE HUMAN SATELLITE DNA SEQUENCES, USUALLY LOCATED NEAR CENTROMERES. CEP SIGNALS ENABLE THE IDENTIFICATION AND ENUMERATION OF HUMAN CHROMOSOMES IN INTERPHASE AND METAPHASE CELLS FROM FRESH AND ARCHIVED SAMPLES. THE LABEL FOR CEP 8 SG ASR PROBE KIT INDICATES IT IS AN ANALYTE SPECIFIC REAGENT, AND ANALYTICAL AND PERFORMANCE CHARACTERISTICS ARE NOT ESTABLISHED. ISSUE: A CUSTOMER RECEIVED SEVEN KITS OF CEP 8 SG ASR PROBE (LIST 06J37-018, LOT 440023), CONTAINING VYSIS CEP 4 (ALPHA SATELLITE) SG PROBE, 20UL, (PART 32-112004 LOT 438814) INSTEAD OF THE EXPECTED CEP 8 PROBE. THE CUSTOMER DID NOT USE THE KITS CONTAINING THE INCORRECT COMPONENT. ABBOTT MOLECULAR SENT SEVEN REPLACEMENT KITS TO THE CUSTOMER. THE CUSTOMER RETURNED THE SEVEN KITS CONTAINING THE INCORRECT COMPONENT TO ABBOTT MOLECULAR. IMPACT: PRODUCT FUNCTIONALITY IS IMPACTED BECAUSE USE OF CEP 4 (SG) WOULD NOT ACCURATELY ENUMERATE CHROMOSOME 8. THE CEP 8 SG PROBE AND THE CEP 4 SG PROBE BOTH PRODUCE GREEN SIGNALS NEAR THE CENTROMERE OF THEIR RESPECTIVE CHROMOSOMES. THE PRESENCE OF SPECTRUM GREEN SIGNALS NEAR THE CENTROMERE MAY LEAD THE TECHNICIAN TO CONCLUDE THE TEST WAS PERFORMED CORRECTLY. IF CEP 8 POSITIVE CONTROLS ARE NOT RUN WITH PATIENT SAMPLES, THEN THE POTENTIAL EXISTS TO GENERATE INCORRECT RESULTS. NOTE: THE CEP 4 PROBE VIALS WERE CORRECTLY LABELED TO IDENTIFY THE ACTUAL CONTENTS OF CEP 4 PROBE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
121783 CEP 8 (D8Z2) SPECTRUMGREEN PROBE ANALYTE SPECIFIC REAGENTS MVU ABBOTT MOLECULAR, INC. 440023

Patients

Seq Age Sex Outcome Treatment
1 Other