FDA Adverse Event
Malfunction
Summary report: N
APOLLO
MDR report key: 3017893
·
Received March 13, 2013
Report
- Report Number
- 9611500-2013-00021
- Event Type
- Malfunction
- Date Received
- March 13, 2013
- Date of Event
- February 8, 2013
- Report Date
- March 13, 2013
- Manufacturer
- DRAGER MEDICAL GMBH
- Product Code
- BSZ
- PMA / PMN Number
- K042607
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Additional Manufacturer Narrative · 1
WE RECEIVED AFFECTED PARTS. THE INVESTIGATION IS ONGOING. WE WILL PROVIDE RESULTS IN A SEPARATE FOLLOW-UP REPORT.
Description of Event or Problem · 1
IT WAS REPORTED THAT: AT THE END OF A CASE, THE ANESTHESIOLOGIST HEARD A DISTINCT POP AND REPORTEDLY THERE WAS A FLAME AROUND THE EXPIRATORY PORT OF THE BREATHING SYSTEM. THE PT WAS DISCONNECTED AND THERE WAS NO INJURY REPORTED. A BIOMED TECH PERFORMED A PRELIMINARY INSPECTION OF THE MACHINE. WHEN HE TOOK THE BREATHING SYSTEM COVER OFF, HE NOTICED EVIDENCE OF CHARRING AND WAS THAT THE CERAMIC DISCS IN SOME OF THE VALVES ON THE VALVE PLATE WERE BROKEN IN PIECES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 106770 | APOLLO | ANESTHESIA MACHINE | BSZ | DRAGER MEDICAL GMBH | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |