UROLUME ENDOPROSTHESIS
Report
- Report Number
- 2183959-2013-00755
- Event Type
- Injury
- Date Received
- March 15, 2013
- Report Date
- March 4, 2013
- Manufacturer
- AMERICAN MEDICAL SYSTEMS, INC.
- Product Code
- MES
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
Narratives
UROLOGY ANNALS, UROL ANN, 2012 SEP-DEC; 4(3): 191-194, FRANCISCO BOTHELHO, ANIL A. THOMAS, RANKO MIOCINOVIC, AND KENNETH W. ANEMEIER, DEPT OF UROLOGY, S. JOAO HOSPITAL, PORTO, PORTUGAL. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE REGARDING THIS EVENT, IT WILL BE RE-EVALUATED AND A FOLLOW-UP REPORT WILL BE SENT.
IT WAS REPORTED THAT A (B)(6) MALE WITH A HISTORY OF NEUROGENIC BLADDER HAD PRESENTING SYMPTOMS OF "RECURRENT URINARY TRACT INFECTIONS, OBSTRUCTIVE VOIDING SYMPTOMS, AND AN INABILITY TO PERFORM SELF-CATHETERIZATION." IT WAS ALSO REPORTED THAT THE PT HAD PREVIOUSLY SUSTAINED A T6 SPINAL CORD INJURY AND HAD A SPHINCTERECTOMY AND UROLUME (URETHRAL STENT) PLACED 10 YEARS AGO TO TREAT DETRUSOR-SPHINCTER DYSSYNERGIA AND BLADDER OUTLET OBSTRUCTION. THE PHYSICIAN FOUND A "CALCIFIED URETHRAL STENT" ON A PLAIN-KIDNEY-URETER-BLADDER RADIOGRAPH. A "FLEXIBLE URETHROSCOPY CONFIRMED THE PRESENCE OF SEVERELY ENCRUSTED STENT AT THE BLADDER NECK, WHICH ALMOST COMPLETELY OBSTRUCTED THE URETHRAL LUMEN." THE PT UNDERWENT LASER FRAGMENTATION AND EXTRACTION OF THE CALCIFIED STENT. UPON COMPLETION OF THE PROCEDURE, THE BLADDER OUTLET WAS NOTED TO BE "PATENT AND WITHOUT ANY EVIDENCE OF RESIDUAL STENT FRAGMENTS." IT WAS ALSO REPORTED THAT THE PT'S "PROSTATIC URETHRA REMAINED AT 7 MONTHS FOLLOW-UP AND THE PT RESUMED INTERMITTENT SELF-CATHETERIZATION WITHOUT DIFFICULTY."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 110564 | UROLUME ENDOPROSTHESIS | ENDOURETHRAL PROSTHESIS | MES | AMERICAN MEDICAL SYSTEMS, INC. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 72 YR | Hospitalization| R |