FDA Adverse Event Malfunction Summary report: N

ADULT DUAL HEATED EVAQUA BREATHING CIRCUIT

MDR report key: 3016284 · Received March 21, 2013

Report

Report Number
9611451-2013-00211
Event Type
Malfunction
Date Received
March 21, 2013
Date of Event
January 21, 2013
Report Date
February 13, 2013
Manufacturer
FISHER & PAYKEL HEALTHCARE LTD
Product Code
BZE
PMA / PMN Number
K983112
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). METHOD: THE COMPLAINT DEVICE WAS RECEIVED AND WAS VISUALLY INSPECTED FOR DAMAGE. RESULTS: VISUAL INSPECTION OF THE EVAQUA EXPIRATORY LIMB REVEALED A HOLE ABOUT 38CM FROM THE PROXIMAL CONNECTOR. THE EVAQUA LIMB APPEARED TO HAVE BEEN PUNCTURED OR SCRATCHED WITH A BLUNT OBJECT. A LOT CHECK REVEALED NO OTHER COMPLAINTS FOR LOT 120827. CONCLUSION: WE WERE UNABLE TO DETERMINE HOW THE LIMB CAME TO BE DAMAGED. (B)(4). THE USER INSTRUCTIONS SUPPLIED WITH THE RT340 BREATHING CIRCUIT STATE: PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT; SET APPROPRIATE VENTILATOR ALARMS; FIT ONLY THE SUPPLIED FISHER & PAYKEL HEALTHCARE CIRCUIT HANGER WITH CARE TO AVOID CIRCUIT DAMAGE.

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED VIA OUR DISTRIBUTOR THAT THE EXPIRATORY LIMB OF AN RT340 ADULT BREATHING CIRCUIT WAS FAULTY. THIS WAS FOUND BEFORE USE ON A PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
117072 ADULT DUAL HEATED EVAQUA BREATHING CIRCUIT BZE BZE FISHER & PAYKEL HEALTHCARE LTD RT340 120827

Patients

Seq Age Sex Outcome Treatment
1