FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 3014303 · Received March 6, 2013

Report

Report Number
1720753-2013-03438
Event Type
Malfunction
Date Received
March 6, 2013
Date of Event
February 25, 2013
Report Date
March 6, 2013
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE SYSTEM BATTERIES WERE REPLACED. THE SYSTEM WAS THEN FOUND TO BE OPERATING AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT DURING A PROCEDURE, THEY HAD TO REBOOT THE SYSTEM. A CHARGER FAILED ERROR MESSAGE HAD OCCURRED. THERE IS NO REPORT OF PATIENT INJURY ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
96539 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1