OT PING METER
Report
- Report Number
- 3008382007-2013-05708
- Event Type
- Malfunction
- Date Received
- March 21, 2013
- Report Date
- February 21, 2013
- Manufacturer
- LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
- Product Code
- NBW
- PMA / PMN Number
- K082590
- Removal / Correction Number
- 2013-IE-005
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
FOLLOW UP #1 (05/20/2013)-DEVICE EVALUATION: THE TEST STRIPS INVOLVED WITH THIS COMPLAINT HAVE BEEN RETURNED TO LFS AND EVALUATED BY PRODUCT ANALYSIS (PA) ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: THE TEST STRIPS PASSED ALL TESTING WITH NO FAULTS FOUND. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, ANOTHER FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.
THE SUBJECT CONTROL SOLUTION AND TEST STRIPS HAVE BEEN RETURNED TO LIFESCAN FOR EVALUATION. THE EVALUATION OF THE PRODUCTS HAVE NOT BEEN COMPLETED; THEREFORE, NO CONCLUSIONS CAN BE DRAWN AT THIS TIME. ONCE THE EVALUATION IS COMPLETED, A SUPPLEMENTAL REPORT WILL BE FILLED.
ON (B)(6) 2013, THE REPORTER CONTACTED LIFESCAN USA, ALLEGING INACCURATE RESULTS WITH CONTROL SOLUTION WITH READINGS OF "111 AND 112MG/DL". THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED RESULTS DID NOT MEET LIFESCAN'S ACCURACY/PRECISION CRITERIA AND THE ALLEGED INACCURACY ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 117753 | OT PING METER | GLUCOSE MONITORING SYS/KIT | NBW | LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL | 3400906 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |