FDA Adverse Event Malfunction Summary report: N

RESERVOIR 1.8ML

MDR report key: 3013869 · Received March 20, 2013

Report

Report Number
3004209178-2013-91999
Event Type
Malfunction
Date Received
March 20, 2013
Date of Event
March 11, 2013
Report Date
March 11, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
FRN
PMA / PMN Number
K001828
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI
Reporter Occupation
PATIENT FAMILY MEMBER OR FRIEND

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED.

Additional Manufacturer Narrative · 1

INSPECTED ONE OPENED AND USED RESERVOIR. PERFORMED MANUAL PRIME AND HIGH PRESSURE TEST PER SPECIFICATIONS. RAN PRIMING IN INSULIN PUMP. RESERVOIR PASSED PER SPECIFICATION. NO LEAKAGE ANOMALY WAS OBSERVED DURING ANALYSIS. CHECKED O-RINGS FOR DEFECTS AND NONE WERE FOUND. RESERVOIR CONNECTED AND LOCKED IN PLACE PROPERLY.

Description of Event or Problem · 1

IT WAS REPORTED THAT INSULIN FROM THE RESERVOIRS LEAKED PAST THE O-RINGS AND INTO THE RESERVOIR COMPARTMENT. THE CUSTOMER'S BLOOD GLUCOSE WAS 422MG/DL. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
115676 RESERVOIR 1.8ML RESERVOIR FRN MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-326A H8489386

Patients

Seq Age Sex Outcome Treatment
1 11 YR