FDA Adverse Event Malfunction Summary report: N

COYOTE¿ ES

MDR report key: 3013451 · Received March 20, 2013

Report

Report Number
2134265-2013-01679
Event Type
Malfunction
Date Received
March 20, 2013
Date of Event
February 25, 2013
Report Date
February 25, 2013
Manufacturer
BOSTON SCIENTIFIC - MAPLE GROVE
Product Code
LIT
PMA / PMN Number
K080982
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4) DEVICE EVALUATED BY MFR.: THE DEVICE WAS NOT RETURNED FOR EVALUATION; THEREFORE A FAILURE ANALYSIS OF THE COMPLAINT DEVICE COULD NOT BE COMPLETED. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL/PROCEDURAL FACTORS. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PERCUTANEOUS PERIPHERAL INTERVENTION A BALLOON RUPTURE OCCURRED. THE 99% STENOSED TARGET LESION WAS LOCATED IN THE CALCIFIED AND MODERATELY TORTUOUS LEFT ANTERIOR TIBIAL ARTERY. THE 1.5 X 20MM X 143CM COYETE ES MR BALLOON CATHETER WAS ADVANCED OVER A NON BSC GUIDE WIRE AND THROUGH A 4.5FR. NON BSC INTRODUCER SHEATH. THE BALLOON WAS INFLATED WITH AN ENCORE INFLATION DEVICE AT 12ATMS AND RUPTURED ON THE 1ST INFLATION. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS IS GOOD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
116463 COYOTE¿ ES CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL LIT BOSTON SCIENTIFIC - MAPLE GROVE H74939135152010 14513290

Patients

Seq Age Sex Outcome Treatment
1