FDA Adverse Event Malfunction Summary report: N

ACCU-CHEK SPIRIT COMBO

MDR report key: 3013366 · Received March 20, 2013

Report

Report Number
2183996-2013-00462
Event Type
Malfunction
Date Received
March 20, 2013
Date of Event
March 9, 2013
Report Date
August 1, 2013
Manufacturer
ROCHE DIABETES CARE AG
Product Code
LZG
PMA / PMN Number
NA
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

THIS INCIDENT OCCURRED OUTSIDE THE UNITED STATES. INFORMATION CONTAINED WITHIN THIS REPORT IS ALL THAT IS AVAILABLE AT THIS TIME. IF FURTHER INFORMATION IS OBTAINED, IT WILL BE PROVIDED IN A SUPPLEMENTAL REPORT.

Additional Manufacturer Narrative · 1

CONCLUSION THE COMPLAINT CANNOT BE VERIFIED, THE PRODUCTION DATA SHOWS NO DEVIATIONS OF THE SPECIFICATIONS. RESULT AS THE PRODUCT HAS NOT BEEN RETURNED FOR INVESTIGATION AND THE PRODUCTION DATA COMPLY WITH THE SPECIFICATION, THE COMPLAINT COULD NOT BE REPLICATED. PRODUCTION REPORTS WERE REVIEWED. AS THE PRODUCT HAS NOT RETURNED FOR EVALUATION, THE COMPLAINT COULD NOT BE INVESTIGATED. DEVICE WAS NOT RETURNED.

Description of Event or Problem · 1

PATIENT REPORTED THE INFUSION DEVICE HAD ACCIDENTAL WATER CONTACT ON (B)(6) 2013. PATIENT STATED NOW THE UP AND DOWN BUTTONS DON'T WORK PROPERLY. PATIENT REPORTED EXPERIENCING NO HEALTH ISSUES. NO FURTHER INFORMATION IS AVAILABLE. NO PHYSIOLOGICAL EFFECTS WERE REPORTED. THE PATIENT DID NOT REQUIRE ASSISTANCE FROM A HEALTHCARE PROFESSIONAL OR SECOND PARTY TO ADDRESS THE ISSUE. REQUESTED RETURN OF THE ALLEGED INFUSION DEVICE FOR EVALUATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
115035 ACCU-CHEK SPIRIT COMBO INSULIN INFUSION PUMP LZG ROCHE DIABETES CARE AG 00700006863 NA

Patients

Seq Age Sex Outcome Treatment
1 032 YR