FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3013250 · Received March 20, 2013

Report

Report Number
2531779-2013-02856
Event Type
Malfunction
Date Received
March 20, 2013
Report Date
February 26, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

FOLLOW-UP #1 (B)(4) - DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: A REVIEW OF THE BLACK BOX SHOWS REBOOTING HAD OCCURRED. THERE WAS NO PHYSICAL DAMAGE OBSERVED TO THE BATTERY COMPARTMENT. THE BATTERY CAP WAS NOT RETURNED WITH THE PUMP FOR INVESTIGATION. A TEST CAP WAS USED TO COMPLETE INVESTIGATION. THE TEST CAP FULLY TIGHTENS TO THE PUMP WITH NO YELLOW O-RING SHOWING. THE PUMP POWERED ON APPROPRIATELY WITH FUNCTIONAL AUDITORY AND VIBRATORY FEATURES. THE PUMP WAS EXERCISED FOR 24 HOURS WITH NO POWER ISSUES OCCURRED. THE PUMP COVER WAS REMOVED AND THERE WAS NO DAMAGE OBSERVED TO THE POWER CIRCUIT. THE COMPLAINT COULD NOT BE DUPLICATED ON INVESTIGATION.

Description of Event or Problem · 1

ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING A POWER (NO POWER) ISSUE. THE REPORTER STATED THAT THE PUMP WAS NOT POWERING ON AFTER MULTIPLE BATTERY CHANGES. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
115864 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 43 YR