FDA Adverse Event Malfunction Summary report: N

XIENCE PRO EVEROLIMUS ELUTING RX CORONARY STENT SYSTEM

MDR report key: 3013246 · Received March 20, 2013

Report

Report Number
2024168-2013-01592
Event Type
Malfunction
Date Received
March 20, 2013
Date of Event
February 13, 2013
Report Date
February 25, 2013
Manufacturer
AV-TEMECULA-CT
Product Code
NIQ
PMA / PMN Number
P070015
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. THE XIENCE PRO IS CURRENTLY NOT COMMERCIALLY AVAILABLE IN THE US. THE DEVICE IS EXPECTED TO BE RETURNED FOR EVALUATION. IT HAS NOT YET BEEN RECEIVED. A FOLLOW-UP REPORT WILL BE SUBMITTED WITH ALL ADDITIONAL RELEVANT INFORMATION.

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED FOR EVALUATION. THE FAILURE TO DEPLOY THE STENT WAS CONFIRMED. BASED ON VISUAL AND FUNCTIONAL ANALYSIS OF THE RETURNED DEVICE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THE REPORTED EVENT. THE RESULTS OF THE QUERY OF SIMILAR INCIDENTS IN THE COMPLAINT HANDLING DATABASE FOR THIS LOT DID NOT INDICATE A MANUFACTURING ISSUE. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A MODERATELY TORTUOUS, MODERATELY CALCIFIED LESION LOCATED IN THE CIRCUMFLEX. AFTER PREDILATATION WAS PERFORMED, THE 3.0X18 MM XIENCE PRO STENT DELIVERY SYSTEM (SDS) WAS ADVANCED TO THE LESION AND THE SDS BALLOON WAS INFLATED 2 TIMES AT 16 ATMOSPHERES FOR STENT DEPLOYMENT; HOWEVER, WHEN THE SDS WAS BEING RETRIEVED FROM THE ANATOMY, IT WAS OBSERVED THAT THE STENT IMPLANT WAS STILL ON THE BALLOON. A NEW XIENCE PRO SDS WAS USED SUCCESSFULLY TO COMPLETE THE CASE. NO ADVERSE PATIENT EFFECTS OR CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE WAS REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
116297 XIENCE PRO EVEROLIMUS ELUTING RX CORONARY STENT SYSTEM DRUG ELUTING CORONARY STENT SYSTEM NIQ AV-TEMECULA-CT 2030241

Patients

Seq Age Sex Outcome Treatment
1