- 1416980-2013-06719
- Event Type
- Malfunction
- Date Received
- March 20, 2013
- Date of Event
- February 1, 2013
- Report Date
- February 27, 2013
- Manufacturer
- BAXTER HEALTHCARE - MALTA
- Product Code
- FPA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BE
- Reporter Occupation
- PHARMACIST
Narratives
(B)(4). THE SAMPLE IS REPORTED TO BE AVAILABLE FOR EVALUATION. IF THE SAMPLE IS RECEIVED OR ADDITIONAL RELEVANT INFORMATION BECOMES AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED.
(B)(4). EVALUATION SUMMARY: THE USED SAMPLE WAS RECEIVED FOR EVALUATION. VISUAL INSPECTION REVEALED THAT THE TUBING WAS BENT JUST UNDER THE VENTED CHAMBER. FUNCTIONAL TESTS REVEALED A RESTRICTED FLOW AND TRACES OF SOLVENT. THE REPORTED CONDITION WAS CONFIRMED. THE ROOT CAUSE WAS NOT DETERMINED. A BATCH REVIEW WAS CONDUCTED AND THERE WERE NO DEVIATIONS FOUND RELATED TO THIS REPORTED CONDITION DURING THE MANUFACTURE OF THIS LOT. SHOULD ADDITIONAL RELEVANT INFORMATION BE RECEIVED, A FOLLOW-UP MEDWATCH WILL BE SUBMITTED.
IT WAS REPORTED TO BAXTER (B)(4) THAT AN IV SOL ADMINISTRATION SET Y 3V LL HAD BLOCKED TUBING JUST UNDER THE DRIP CHAMBER. THE REPORTED CONDITION OCCURRED DURING PRIMING. THERE IS NO REPORT OF PATIENT INVOLVEMENT, INJURY/ADVERSE EVENTS, OR MEDICAL INTERVENTION IN ASSOCIATION WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.