Manufacturer
BAXTER HEALTHCARE - MALTA
Product Code
FPA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BE
Reporter Occupation
PHARMACIST

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE SAMPLE IS REPORTED TO BE AVAILABLE FOR EVALUATION. IF THE SAMPLE IS RECEIVED OR ADDITIONAL RELEVANT INFORMATION BECOMES AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE USED SAMPLE WAS RECEIVED FOR EVALUATION. VISUAL INSPECTION REVEALED THAT THE TUBING WAS BENT JUST UNDER THE VENTED CHAMBER. FUNCTIONAL TESTS REVEALED A RESTRICTED FLOW AND TRACES OF SOLVENT. THE REPORTED CONDITION WAS CONFIRMED. THE ROOT CAUSE WAS NOT DETERMINED. A BATCH REVIEW WAS CONDUCTED AND THERE WERE NO DEVIATIONS FOUND RELATED TO THIS REPORTED CONDITION DURING THE MANUFACTURE OF THIS LOT. SHOULD ADDITIONAL RELEVANT INFORMATION BE RECEIVED, A FOLLOW-UP MEDWATCH WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED TO BAXTER (B)(4) THAT AN IV SOL ADMINISTRATION SET Y 3V LL HAD BLOCKED TUBING JUST UNDER THE DRIP CHAMBER. THE REPORTED CONDITION OCCURRED DURING PRIMING. THERE IS NO REPORT OF PATIENT INVOLVEMENT, INJURY/ADVERSE EVENTS, OR MEDICAL INTERVENTION IN ASSOCIATION WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
116291 ACCESS SET, ADMINISTRATION, INTRAVASCULAR FPA BAXTER HEALTHCARE - MALTA 12D06V069