SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Report
- Report Number
- 1416980-2013-06715
- Event Type
- Malfunction
- Date Received
- March 20, 2013
- Date of Event
- March 4, 2013
- Report Date
- March 4, 2013
- Manufacturer
- BAXTER HEALTHCARE
- Product Code
- FKX
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- PATIENT
Narratives
(B)(4). THIS COMPLAINT WAS NOT CONFIRMED AND THE ROOT CAUSE COULD NOT BE DETERMINED BECAUSE THE DISPOSABLE CASSETTE WAS NOT RETURNED TO BAXTER FOR EVALUATION, A LOT NUMBER WAS NOT PROVIDED FOR A BATCH REVIEW AND THE USER DID NOT DESCRIBE ANY TYPES OF USE ERRORS OR OTHER ISSUES THAT MIGHT HAVE CAUSED THE REPORTED EVENT.
IT WAS REPORTED THAT A PATIENT HAD A SYSTEM ERROR (SE) 2240 OCCUR DURING PERITONEAL DIALYSIS (PD) THERAPY WHILE USING A HOMECHOICE IN CONJUNCTION WITH A DISPOSABLE CASSETTE. THE HOME PATIENT (HP) WAS CONNECTED AND IN DWELL 4 OF 6. THE HP STATED THAT SOLUTION WAS IN HEATER BAG ONLY, THE SUPPLY BAG IS EMPTY, AND THERE WAS NO DISCONNECTION FROM THE CASSETTE. THE PATIENT LINE WAS PROPERLY PRIMED BEFORE CONNECTING, NO PATIENT EXTENSION LINES WERE USED, THE HP DID NOT DISCONNECT PRIOR TO THE ALARM, ALL BAGS WERE PROPERLY CONNECTED, THERE WERE NO OPEN CLAMPS ON ANY UNUSED SUPPLY LINES, AND THERE WAS NOTHING UNUSUAL NOTED ABOUT THE SUPPLIES. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) REVIEWED THE ALARM WITH THE HP AND ASSISTED THEM WITH TROUBLESHOOTING AND CLEARING THE ALARM. PROPER PROCEDURES PER THE USER MANUAL WERE REVIEWED WITH THE HP. THERE WAS PATIENT INVOLVEMENT BUT THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION INDICATED AT THE TIME OF THE INITIAL REPORT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 116075 | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | HOMECHOICE |