FDA Adverse Event Malfunction Summary report: N

ETRIO

MDR report key: 3013147 · Received March 20, 2013

Report

Report Number
3005075853-2013-01321
Event Type
Malfunction
Date Received
March 20, 2013
Date of Event
February 7, 2013
Report Date
March 14, 2013
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GEI
PMA / PMN Number
K072177
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME WHEN ADDITIONAL INFORMATION IS RECEIVED AND/OR THE DEVICE ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL MEDWATCH WILL BE SENT.

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RECEIVED WITH THE ELECTRODE SEPARATED FROM THE CERAMIC AND THE ELECTRODE IS STILL ATTACHED TO THE ACTIVE ROD. THE CERAMIC IS DAMAGED AND PARTIALLY MISSING. THE CERAMIC WAS DAMAGED BY THE SEPARATION OF THE ELECTRODE FROM THE LOWER JAW. TESTING FOUND A SHORT ON THE DEVICE WHEN THE JAWS ARE FULLY OR PARTIALLY CLOSED, RESULTING IN THE "REPOSITION JAWS AND REACTIVE" YELLOW MESSAGE SCREEN IS ADVISING THAT THE INSTRUMENT IS BEING ACTIVATED ON LOW IMPEDANCE (THIN) TISSUE OR METAL (SUCH AS STAPLES, CLIPS, RETRACTORS, OR CLAMPS). THIS ERROR SCREEN CAN ALSO BE A RESULT OF THE ELECTRODE SHORTING WITH THE UPPER JAW.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LAPAROSCOPIC HYSTERECTOMY PROCEDURE, THE ELECTRODE SEPARATED FROM THE JAW DURING USE. UNKNOWN HOW CASE WAS COMPLETED. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
116504 ETRIO ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES GEI ETHICON ENDO-SURGERY, LLC. UNK J4C22L

Patients

Seq Age Sex Outcome Treatment
1 GENERATOR