ETRIO
Report
- Report Number
- 3005075853-2013-01321
- Event Type
- Malfunction
- Date Received
- March 20, 2013
- Date of Event
- February 7, 2013
- Report Date
- March 14, 2013
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GEI
- PMA / PMN Number
- K072177
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SZ
- Reporter Occupation
- OTHER
Narratives
(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME WHEN ADDITIONAL INFORMATION IS RECEIVED AND/OR THE DEVICE ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL MEDWATCH WILL BE SENT.
(B)(4). THE DEVICE WAS RECEIVED WITH THE ELECTRODE SEPARATED FROM THE CERAMIC AND THE ELECTRODE IS STILL ATTACHED TO THE ACTIVE ROD. THE CERAMIC IS DAMAGED AND PARTIALLY MISSING. THE CERAMIC WAS DAMAGED BY THE SEPARATION OF THE ELECTRODE FROM THE LOWER JAW. TESTING FOUND A SHORT ON THE DEVICE WHEN THE JAWS ARE FULLY OR PARTIALLY CLOSED, RESULTING IN THE "REPOSITION JAWS AND REACTIVE" YELLOW MESSAGE SCREEN IS ADVISING THAT THE INSTRUMENT IS BEING ACTIVATED ON LOW IMPEDANCE (THIN) TISSUE OR METAL (SUCH AS STAPLES, CLIPS, RETRACTORS, OR CLAMPS). THIS ERROR SCREEN CAN ALSO BE A RESULT OF THE ELECTRODE SHORTING WITH THE UPPER JAW.
IT WAS REPORTED THAT DURING A LAPAROSCOPIC HYSTERECTOMY PROCEDURE, THE ELECTRODE SEPARATED FROM THE JAW DURING USE. UNKNOWN HOW CASE WAS COMPLETED. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 116504 | ETRIO | ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES | GEI | ETHICON ENDO-SURGERY, LLC. | UNK | J4C22L |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | GENERATOR |