FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 3013062 · Received March 1, 2013

Report

Report Number
1720753-2013-03254
Event Type
Malfunction
Date Received
March 1, 2013
Date of Event
February 15, 2013
Report Date
March 1, 2013
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE REP EVALUATED THE SYSTEM AND REPLACED THE 5 VOLT POWER SUPPLY AND THE CONTROL PANEL PROCESSOR BOARD. THE SYSTEM OPERATES AS INTENDED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE SYSTEM DISPLAYED A CONTROL PANEL COMMUNICATION ERROR. THIS ERROR WILL CAUSE THE SYSTEM TO LOCKUP, SHUT DOWN, OR NOT TO BOOT. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
90459 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1