FDA Adverse Event Malfunction Summary report: N

TERUMO TCM II COOLING AND HEATING SYSTEM

MDR report key: 3012961 · Received March 1, 2013

Report

Report Number
1828100-2013-00199
Event Type
Malfunction
Date Received
March 1, 2013
Date of Event
February 8, 2013
Report Date
February 8, 2013
Manufacturer
TERUMO CARDIOVASCULAR SYSTEM CORP.
Product Code
DWC
PMA / PMN Number
K883603
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE FIELD SERVICE REPRESENTATIVE (FSR) CONFIRMED THE COMPLAINT. THE ICE BLOCK HAS FROZEN NEARLY 90% OF TANK AND WAS FROZEN SOLID THE ENTIRE DEPTH OF THE FRONT HALF. FLOW WAS WORKING, BUT VOLUME AVAILABLE WAS MINIMAL. THE FSR DEFROSTED THE UNIT. FLOWS WERE STILL WEAK AND THERE WAS NOTICEABLE SEDIMENT IN BOTH TANKS. THE FSR DRAINED UNIT AND OPENED BOTTOM LINES; THE VALVES AND LINES WERE CLEAR. THE FSR REPLACED THE PRESSURE SWITCH BECAUSE IT WAS NOT OPERATING. SEE RELATED MDR #1828100-2013-00220. HE REPLACED VALVES 1 AND 2 ON THE UNIT. WITH THE COMPONENTS REPLACED AND A FULL INSPECTION PERFORMED, THE UNIT OPERATES TO MANUFACTURER SPECIFICATIONS AND WAS RETURNED TO CLINICAL USE. IF ADDITIONAL INFORMATION BECOMES AVAILABLE ON THIS COMPLAINT THAT WOULD ALTER THE FACTS AND/OR CONCLUSION, A SUPPLEMENTAL REPORT WILL BE FILED ACCORDINGLY.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING USE OF THE DEVICE FOR A CARDIOPULMONARY BYPASS PROCEDURE, THE CARDIOPLEGIA WOULD NOT COOL AND THERE WAS LITTLE FLOW. THE PERFUSIONIST ALSO NOTED THAT THE LARGE ICE TANK WAS FROZEN SOLID. AS A RESULT, AN ALTERNATE DEVICE WAS EMPLOYED. THE SURGICAL PROCEDURE WAS COMPLETED SUCCESSFULLY, AND THERE WERE NO DELAYS, NO BLOOD LOSS OR NO ADVERSE CONSEQUENCES TO THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
90351 TERUMO TCM II COOLING AND HEATING SYSTEM DWC TERUMO CARDIOVASCULAR SYSTEM CORP. 4415

Patients

Seq Age Sex Outcome Treatment
1