LIFEVEST WCD 4000 SYSTEM
MDR report key: 3012953
·
Received February 26, 2013
Report
- Report Number
- 3008642652-2013-00547
- Event Type
- Malfunction
- Date Received
- February 26, 2013
- Date of Event
- February 8, 2013
- Report Date
- February 20, 2013
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVAL OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (SERVICE CODE 204) WAS CONFIRMED. UPON INVESTIGATION, THE TRUNK CABLE CONNECTOR WAS DAMAGED AND THE ORANGE (+5V) AND BLACK (MAIN BATT) WIRES WERE OPEN INSIDE OF THE CONNECTOR. THE ROOT CAUSE FOR THE BROKEN CONNECTOR WIRES COULD NOT BE POSITIVELY IDENTIFIED BUT WAS LIKELY EXCESSIVE FORCE. NO ADVERSE EVENT RESULTED FROM THE OPEN CONNECTOR WIRES. THE PT RECEIVED A REPLACEMENT ELECTRODE BELT.
Description of Event or Problem · 1
A (B)(6) MALE PT CALLED ZOLL CUSTOMER SUPPORT TO REPORT THAT HE WAS RECEIVING A SERVICE CODE 204. THE PT WAS ISSUED A REPLACEMENT ELECTRODE BELT.