FDA Adverse Event
Malfunction
Summary report: N
LIFEVEST WCD 4000 SYSTEM
MDR report key: 3012949
·
Received February 26, 2013
Report
- Report Number
- 3008642652-2013-00498
- Event Type
- Malfunction
- Date Received
- February 26, 2013
- Date of Event
- January 27, 2013
- Report Date
- February 15, 2013
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVAL OF MONITOR SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (CHECK ELECTRODE BELT MESSAGES) WAS CONFIRMED. UPON INVESTIGATION, THE MONITOR'S ELECTRODE BELT CONNECTOR WAS BROKEN FREE FROM THE MONITOR CASE AND THE BLACK PULSE WIRE WAS BROKEN. THE ROOT CAUSE FOR THE DAMAGED CONNECTOR COULD NOT BE POSITIVELY IDENTIFIED, BUT THE DAMAGE LIKELY RESULTED FROM THE MONITOR BEING DROPPED WHILE CONNECTED TO AN ELECTRODE BELT. NO ADVERSE EVENT RESULTED FROM THE DAMAGED MONITOR CONNECTOR. THE PT RECEIVED A REPLACEMENT MONITOR.
Description of Event or Problem · 1
A (B)(6) MALE PT CALLED ZOLL CUSTOMER SUPPORT TO REPORT THAT HE WAS RECEIVING CHECK ELECTRODE BELT MESSAGES. THE PT WAS ISSUED A REPLACEMENT MONITOR.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 83192 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 4000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 63 YR |