FDA Adverse Event
Malfunction
Summary report: N
840 VENTILATOR
MDR report key: 3012878
·
Received March 1, 2013
Report
- Report Number
- 8020893-2013-00473
- Event Type
- Malfunction
- Date Received
- March 1, 2013
- Date of Event
- February 4, 2013
- Report Date
- February 4, 2013
- Manufacturer
- COVIDIEN, FORMERLY NELLCO
- Product Code
- CBK
- PMA / PMN Number
- K970460
- Report Source
- Manufacturer report
- Reporter Location
- SP
- Reporter Occupation
- NOT APPLICABLE
Narratives
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT AN 840 VENTILATOR HAD A BLANK SCREEN. NO PT INVOLVEMENT. THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) VERIFIED THE MALFUNCTION. THE CSE INSPECTED THE DEVICE AND REPLACED THE GRAPHIC USER INTERFACE (GUI) CPU AND BACKLIGHT PCBS. THE UNIT PASSED EXTENDED SELF-TESTING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 90154 | 840 VENTILATOR | VENTITLATOR | CBK | COVIDIEN, FORMERLY NELLCO | 840 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |