FDA Adverse Event
Malfunction
Summary report: N
840 VENTILATOR
MDR report key: 3012832
·
Received March 1, 2013
Report
- Report Number
- 8020893-2013-00513
- Event Type
- Malfunction
- Date Received
- March 1, 2013
- Date of Event
- February 7, 2013
- Report Date
- February 8, 2013
- Manufacturer
- COVIDIEN, FORMERLY NELLCOR
- Product Code
- CBK
- PMA / PMN Number
- K970460
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE 840 VENTILATOR HAD A BLANK DISPLAY WHILE IN USE ON A PATIENT. THE PATIENT WAS NOT HARMED OR INJURED AS A RESULT OF THE EVENT. THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) VERIFIED THE MALFUNCTION. THE CSE INSPECTED THE DEVICE AND REPLACED THE POWER SUPPLY. THE UNIT PASSED EXTENDED SELF-TESTING.