FDA Adverse Event Malfunction Summary report: N

SYSTEM, PERITONEAL, AUTOMATIC DELIVERY

MDR report key: 3012797 · Received March 20, 2013

Report

Report Number
1416980-2013-06685
Event Type
Malfunction
Date Received
March 20, 2013
Date of Event
January 5, 2013
Report Date
February 27, 2013
Manufacturer
BAXTER HEALTHCARE
Product Code
FKX
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MX
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DURING EVALUATION OF A HOMECHOICE HARDWARE DEVICE, THE ALARM WAS CONFIRMED THROUGH THE EVENT HISTORY LOG REVIEW. THIS ALARM IS INDICATIVE OF A POTENTIAL MALFUNCTION OF THE DISPOSABLE CASSETTE. AS THE CASSETTE WAS NOT RETURNED AND THE LOT NUMBER IS UNKNOWN, A DEVICE ANALYSIS OF THE CASSETTE CANNOT BE COMPLETED.

Description of Event or Problem · 1

DURING EVALUATION OF A RETURNED HOMECHOICE DEVICE, A SYSTEM ERROR 2240 ALARM (AIR IN LINE) WAS IDENTIFIED IN THE LOG, WHICH INDICATES A POTENTIAL MALFUNCTION OF THE DISPOSABLE CASSETTE. THE ALARM OCCURRED ON (B)(6) 2013 AT 23:09:03. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
115950 SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE

Patients

Seq Age Sex Outcome Treatment
1 HOMECHOICE