FDA Adverse Event Malfunction Summary report: N

TOTALCARE BED

MDR report key: 3012722 · Received February 22, 2013

Report

Report Number
1824206-2013-01076
Event Type
Malfunction
Date Received
February 22, 2013
Date of Event
January 27, 2013
Report Date
January 27, 2013
Manufacturer
HILL-ROM, INC.
Product Code
FNL
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

DURING AN INSPECTION, THE HILL-ROM TECH FOUND A DAMAGED POWER CORD WITH A BROKEN GROUND PRONG. THE TECH REPLACED THE POWER CORD TO RESOLVE THE ISSUE.

Description of Event or Problem · 1

INFO RECEIVED INDICATED THE BED HAD AN INTERMITTENT GROUND LINE CONNECTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
79054 TOTALCARE BED AC POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM, INC. 1900

Patients

Seq Age Sex Outcome Treatment
1