FDA Adverse Event
Malfunction
Summary report: N
TOTALCARE BED
MDR report key: 3012722
·
Received February 22, 2013
Report
- Report Number
- 1824206-2013-01076
- Event Type
- Malfunction
- Date Received
- February 22, 2013
- Date of Event
- January 27, 2013
- Report Date
- January 27, 2013
- Manufacturer
- HILL-ROM, INC.
- Product Code
- FNL
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
DURING AN INSPECTION, THE HILL-ROM TECH FOUND A DAMAGED POWER CORD WITH A BROKEN GROUND PRONG. THE TECH REPLACED THE POWER CORD TO RESOLVE THE ISSUE.
Description of Event or Problem · 1
INFO RECEIVED INDICATED THE BED HAD AN INTERMITTENT GROUND LINE CONNECTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 79054 | TOTALCARE BED | AC POWERED ADJUSTABLE HOSPITAL BED | FNL | HILL-ROM, INC. | 1900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |