INFUSOR
Report
- Report Number
- 1416980-2013-06669
- Event Type
- Malfunction
- Date Received
- March 20, 2013
- Date of Event
- February 19, 2013
- Report Date
- February 28, 2013
- Manufacturer
- BAXTER HEALTHCARE - IRVINE
- Product Code
- MEB
- PMA / PMN Number
- K011317
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NL
- Reporter Occupation
- PHARMACIST
Narratives
(B)(4). EVALUATION SUMMARY: BAXTER RECEIVED ONE SAMPLE CONTAINING APPROXIMATELY 100 ML OF FLUID IN THE RESERVOIR. VISUAL EXAMINATION OF THE SAMPLE SHOWED NO SIGN OF PHYSICAL ABNORMALITY. AN ACCURACY FLOW TEST WAS PERFORMED ON THE SAMPLE. THE INFUSOR PRODUCED FUNCTIONAL RESULTS WITHIN THE SPECIFICATION RANGE; THE DEVICE PERFORMED AS EXPECTED. THE REPORTED CONDITION OF AN UNDERINFUSION WAS NOT CONFIRMED. NO ROOT CAUSE WAS IDENTIFIED, AS NO EVIDENCE OF PRODUCT MALFUNCTION WAS OBSERVED. NO REPAIR WAS DONE, AS THIS IS A SINGLE-USE DEVICE WHICH WILL BE DISCARDED. NO OTHER OBSERVATIONS WERE NOTED ON THE UNIT. THE BATCH REVIEW REVEALED THAT ALL OF THE ACCEPTANCE CRITERIA WERE MET TO RELEASE THE LOT. THERE WERE NO NON-CONFORMANCES, FAILURES, REWORK, OR DEVIATIONS RELATED TO THE LOT. THERE WERE NO CHANGES TO SPECIFICATIONS, TEST METHODS, PROCESS, EQUIPMENT, OR RAW MATERIALS THAT COULD BE ASSOCIATED WITH THE REPORTED CONDITION.
BAXTER NETHERLANDS RECEIVED A REPORT THAT A MULTIRATE INFUSOR UNDERINFUSED DURING PATIENT USE. THE INFUSOR INFUSED AT 6 ML/HOUR INSTEAD OF THE EXPECTED INFUSION OF 12 ML/HOUR. THE DEVICE WAS FILLED WITH 276 ML OF SOLUTION. THE CONCOMITANT MEDICAL PRODUCTS ARE CURRENTLY UNKNOWN. THERE WAS PATIENT INVOLVEMENT, BUT THERE WAS NO REPORT OF PATIENT INJURY, MEDICAL INTERVENTION NECESSARY, OR ADVERSE REACTION IN ASSOCIATION WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 116071 | INFUSOR | PUMP, INFUSION, ELASTOMERIC | MEB | BAXTER HEALTHCARE - IRVINE | 12J043 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |