FDA Adverse Event Malfunction Summary report: N

OMNIPOD INSULIN PUMP

MDR report key: 3012622 · Received February 21, 2013

Report

Report Number
3004464228-2013-00151
Event Type
Malfunction
Date Received
February 21, 2013
Date of Event
January 25, 2013
Report Date
January 25, 2013
Manufacturer
INSULET CORPORATION
Product Code
LZG
PMA / PMN Number
K042792
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS NOT RETURNED FOR EVAL. WE ARE UNABLE TO CONFIRM THE KINKED, BENT CANNULA, OR DETERMINE IF IT COULD HAVE CONTRIBUTED TO THE REPORTED HYPERGLYCEMIA. QUALIFICATION RECORDS WERE REVIEWED AND THE PRODUCT LOT MET ALL ACCEPTANCE CRITERIA. THE OMNIPOD USER GUIDE WARNS, "TEST RESULTS GREATER THAN 250 MG/DL MEAN HIGH BLOOD GLUCOSE (HYPERGLYCEMIA). IF YOU GET RESULTS ABOVE 250 MG/DL, BUT DO NOT HAVE SYMPTOMS OF HYPERGLYCEMIA, REPEAT THE TEST. IF YOU HAVE SYMPTOMS OR CONTINUE TO GET RESULTS THAT ABOVE 250 MG/DL, FOLLOW THE TREATMENT ADVICE OF YOUR HEALTHCARE PROVIDER".

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT SHE ACTIVATED THE DEVICE THIS MORNING AND AFTER ABOUT THREE HOURS, HER BLOOD GLUCOSE ROSE TO 280 MG/DL. AFTER POD REMOVAL, SHE OBSERVED THAT THE CANNULA WAS BENT AND KINKED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
77405 OMNIPOD INSULIN PUMP PUMP, INFUSION, INSULIN LZG INSULET CORPORATION 11200 L30960

Patients

Seq Age Sex Outcome Treatment
1 40 YR