ACCU-CHEK ® SPIRIT COMBO
Report
- Report Number
- 2183996-2013-00450
- Event Type
- Malfunction
- Date Received
- March 20, 2013
- Date of Event
- March 2, 2013
- Report Date
- May 13, 2013
- Manufacturer
- ROCHE DIABETES CARE AG
- Product Code
- LZG
- PMA / PMN Number
- NA
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
CONCLUSION THE COMPLAINT CANNOT BE VERIFIED DUE TO CARELESS HANDLING OF THE PRODUCT. RESULT THE SOFTCOMPONENTS OF THE UP BUTTON IS LACERATED. THE DAMAGES DID NOT INFLUENCE THE FUNCTIONS OF THE PUMP BUTTONS. THE PROBLEM MENTIONED BY THE CUSTOMER IS RELATED TO CARELESS HANDLING OF THE PRODUCT.
THE INCIDENT OCCURRED OUTSIDE THE UNITED STATES. INFORMATION CONTAINED WITHIN THIS REPORT IS ALL THAT IS AVAILABLE AT THIS TIME. IF FURTHER INFORMATION IS OBTAINED, IT WILL BE PROVIDED IN THE SUPPLEMENTAL REPORT.
PATIENT REPORTED THE BUTTONS ON THE INFUSION DEVICE WORK INTERMITTENTLY. NO PHYSIOLOGICAL EFFECTS WERE REPORTED, AND HE DID NOT REQUIRE TREATMENT FROM A HEALTHCARE REPRESENTATIVE OR SECOND PARTY TO ADDRESS THE ISSUE. THE INFUSION DEVICE WAS REQUESTED FOR EVALUATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 115242 | ACCU-CHEK ® SPIRIT COMBO | INSULIN INFUSION PUMP | LZG | ROCHE DIABETES CARE AG | 00700006863 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 047 YR |