FDA Adverse Event Malfunction Summary report: N

EXETER 2.5 I M PLUG 14MM

MDR report key: 3012584 · Received March 20, 2013

Report

Report Number
0002249697-2013-01041
Event Type
Malfunction
Date Received
March 20, 2013
Date of Event
March 1, 2013
Report Date
March 1, 2013
Manufacturer
STRYKER ORTHOPAEDICS-MAHWAH
Product Code
JDI
PMA / PMN Number
K980843
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NZ
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

WHEN COMPLETED, THE EVALUATION SUMMARY WILL BE SUBMITTED IN A SUPPLEMENTAL REPORT.

Additional Manufacturer Narrative · 1

AN EVENT REGARDING INTRAOPERATIVE FRACTURE INVOLVING A EXETER BONE PLUG WAS REPORTED. THE EVENT WAS CONFIRMED. A MATERIAL ANALYSIS HAS BEEN PERFORMED. THE REPORT CONCLUDED: ¿THE EXETER PLUG FRACTURED IN OVERLOAD, LIKELY DUE TO THE FORCE REQUIRED TO SEAT THE DEVICE. FRACTURE INITIATED ALONG THE PROXIMAL EDGE AND PROPAGATED AXIALLY AROUND THE DEVICE. THERE WAS NO EVIDENCE OF MANUFACTURING OR MATERIAL DEFECTS ON THE DEVICE FEATURED EVALUATED.¿ DEVICE HISTORY REVIEW INDICATED THERE HAVE BEEN NO REPORTED DISCREPANCIES FOR THE REFERENCED LOT. COMPLAINT HISTORY REVIEW INDICATED THERE HAVE BEEN NO OTHER EVENTS FOR THE REPORTED LOT. THE INVESTIGATION CONCLUDED THAT DEVICE FRACTURE WAS CAUSED BY USER MISUSE BY APPLYING TOO MUCH FORCE DURING IMPLANTATION.

Description of Event or Problem · 1

THE INTRAMEDULLARY PLUG WAS INSERTED INTO THE FEMORAL CANAL USING THE PROPER INSERTER AFTER THE CANAL HAD BEEN CORRECTLY SIZED USING THE EXETER SIZERS. THE PLUG SLOWED DOWN AROUND MID-CANAL BUT WITH GENTLE TAPS IT PASSED THROUGH TO THE DESIRED DEPTH. WHEN THE INSERTER WAS REMOVED THE TIP OF THE PLUG WAS FOUND STILL ATTACHED. IT WAS DECIDED NOT TO TRY AND RETRIEVE THE REMAINS OF THE IM PLUG, BUT INSTEAD TO PLACE A ON TOP.

Description of Event or Problem · 1

THE INTRAMEDULLARY PLUG WAS INSERTED INTO THE FEMORAL CANAL USING THE PROPER INSERTER AFTER THE CANAL HAD BEEN CORRECTLY SIZED USING THE EXETER SIZERS. THE PLUG SLOWED DOWN AROUND MID-CANAL BUT WITH GENTLE TAPS IT PASSED THROUGH TO THE DESIRED DEPTH. WHEN THE INSERTER WAS REMOVED THE TIP OF THE PLUG WAS FOUND STILL ATTACHED. IT WAS DECIDED NOT TO TRY AND RETRIEVE THE REMAINS OF THE IM PLUG, BUT INSTEAD TO PLACE A ON TOP.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
116012 EXETER 2.5 I M PLUG 14MM IMPLANT JDI STRYKER ORTHOPAEDICS-MAHWAH L6102

Patients

Seq Age Sex Outcome Treatment
1 59 YR Other