FDA Adverse Event Malfunction Summary report: N

ZENITH FLEX AAA ENDOVASCULAR GRAFT BIFURCATED MAIN BODY

MDR report key: 3012506 · Received February 20, 2013

Report

Report Number
1820334-2013-00092
Event Type
Malfunction
Date Received
February 20, 2013
Date of Event
February 2, 2013
Report Date
February 2, 2013
Manufacturer
COOK, INC.
Product Code
MIH
PMA / PMN Number
P020018
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

CONVERSION TO OPEN SURGICAL REPAIR IS LABELED IN THE IFU. DIFFICULTY IN PLACING DEVICE IS ADDRESSED IN THE IFU. EVENT EVALUATION: STILL UNDER INVESTIGATION.

Description of Event or Problem · 1

STENT GRAFT PLACEMENT WAS CONDUCTED TO A (B)(6) MALE PT WITH RIGHT APPROACH AS AN EMERGENCY PROCEDURE DUE TO RAPTURE OF ABDOMINAL AORTA ON (B)(6) 2013. THE PT WAS SUITABLE FOR ENDOVASCULAR REPAIR THOUGH THERE WAS SEVERE TORTUOSITY AND CALCIFICATION IN THE ACCESS ROUTE AND THE PROCEDURE WAS CONDUCTED AS LABELED. THE DELIVERY SYSTEM OF THE DEVICE DID NOT ADVANCE AT ALL WITH RIGHT APPROACH WITH ULTRA STIFF WIRE GUIDE FIRST. THEN, WIRE GUIDE WAS CHANGED TO LUNDERQUIST AND ADVANCEMENT WAS ATTEMPTED AGAIN WITH SOME FORCE TO PUSH IT. THE DELIVERY SYSTEM COULD ADVANCE SOME, BUT COULD NOT REACH TO THE DESIRED POSITION. THEN APPROACHING SITE WAS CHANGED TO THE LEFT AND ADVANCEMENT OF THE DELIVERY SYSTEM WAS PERFORMED WITH ULTRA STIFF WIRE GUIDE FIRST, BUT BECAUSE IT FAILED, THE WIRE GUIDE WAS CHANGED TO LUNDERQUIST. HOWEVER, IT STUCK AGAIN BEFORE REACHING TO THE DESIRED POSITION AND IT BECAME UNABLE FOR THE PHYSICIAN TO ADVANCE OR REMOVE THE DELIVERY SYSTEM. STENT GRAFT PLACEMENT WAS ABANDONED AND THE PHYSICIAN DECIDED TO CONDUCT AN OPEN SURGERY INSTEAD. AFTER THAT, THE DELIVERY SYSTEM COULD BE REMOVED BY PULLING IT. THERE HAS BEEN NO ADVERSE EFFECTS TO THE PT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
74799 ZENITH FLEX AAA ENDOVASCULAR GRAFT BIFURCATED MAIN BODY MIH SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT MIH COOK, INC. NA 3077987

Patients

Seq Age Sex Outcome Treatment
1 89 YR