FDA Adverse Event Injury Summary report: N

LAMITRODE S8

MDR report key: 3012215 · Received March 12, 2013

Report

Report Number
1627487-2013-10089
Event Type
Injury
Date Received
March 12, 2013
Date of Event
January 25, 2013
Report Date
February 18, 2013
Manufacturer
ST. JUDE MEDICAL - NEUROMODULATION
Product Code
LGW
PMA / PMN Number
P010032
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BR
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

SJM HAS LIMITED INFORMATION RELATED TO THE PT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PT'S PHYSICIAN REGARDING MEDICAL HISTORY.

Description of Event or Problem · 1

IT WAS REPORTED THE PT ((B)(6)) WAS NO LONGER RECEIVING EFFECTIVE STIMULATION. DIAGNOSTIC TESTING IDENTIFIED HIGH IMPEDANCE VALUES ON ALL OF THE SCS LEAD ELECTRODES. THE LEAD WAS EXPLANTED AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
104482 LAMITRODE S8 SCS LEAD LGW ST. JUDE MEDICAL - NEUROMODULATION 3286 3162647

Patients

Seq Age Sex Outcome Treatment
1 UNK Required Intervention