FDA Adverse Event Injury Summary report: N

CHARGING SYSTEM

MDR report key: 3012141 · Received March 13, 2013

Report

Report Number
1627487-2013-06055
Event Type
Injury
Date Received
March 13, 2013
Date of Event
February 20, 2013
Report Date
February 20, 2013
Manufacturer
ST. JUDE MEDICAL - NEUROMODULATION
Product Code
LGW
PMA / PMN Number
P010032
Removal / Correction Number
1627487-07262012-001-C
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THIS CHARGER MODEL WAS ASSOCIATED WITH A FIELD CORRECTION. MFR'S EVAL: CORRECTIVE AND PREVENTIVE ACTION (CAPA) INVESTIGATION WAS PERFORMED. RESULTS: POCKET HEATING WAS CONFIRMED. THE INVESTIGATION FOR CAPA (B)(4) ASSOCIATED WITH HEATING WHILE CHARGING (POCKET HEATING) CONCLUDED THAT THE CHARGER WAS CAPABLE OF TRANSFERRING ENERGY TO THE IPG AT A RATE THAT WOULD CAUSE HEATING OF THE IPG AND/OR CHARGING WAND OF SUFFICIENT ELEVATED TEMPERATURE TO CAUSE PAIN AND BURNS. THE HEATING WHILE CHARGING WAS DETERMINED TO E EXACERBATED BY OFF-AXIS CHARGING OF SHALLOW IMPLANTED IPGS AND THAT ALL CHARGERS WERE CAPABLE OF ELEVATED HEATING. SJM HAS LIMITED INFO RELATED TO THE PT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PT'S PHYSICIAN REGARDING MEDICAL HISTORY.

Description of Event or Problem · 1

DEVICE 4 OF 4. REFERENCE MFR REPORT: 1627487-2013-06052, 06053, 06054. IT WAS REPORTED THE PT EXPERIENCES HEATING WHILE RECHARGING. THE PT HAS TWO SYSTEMS, BUT HE STATED HE ONLY USES ONE CHARGER TO CHARGE BOTH IPGS AND HE EXPERIENCES HEATING WHEN CHARGING BOTH IPGS. ALSO, THE PT REQUESTED ONLY ONE REPLACEMENT CHARGER BE SENT AS HE DOES NOT WANT TO HAVE TWO CHARGERS. A LOW ENERGY CHARGER WAS SENT TO ADDRESS THIS ISSUE. F/U INDICATES THE NEW CHARGER IS WORKING FINE AND THE PT IS NO LONGER EXPERIENCING ANY HEATING WHILE CHARGING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
106491 CHARGING SYSTEM SCS CHARGING SYSTEM LGW ST. JUDE MEDICAL - NEUROMODULATION 3721 3449505

Patients

Seq Age Sex Outcome Treatment
1 75 YR Required Intervention SCS LEAD: MODEL 3286| IMPLANT DATE:| SCS LEAD: MODEL 3228| IMPLANT DATE: