ACCU-CHEK SPIRIT
Report
- Report Number
- 2183996-2013-00441
- Event Type
- Malfunction
- Date Received
- March 20, 2013
- Date of Event
- March 6, 2013
- Report Date
- April 11, 2013
- Manufacturer
- ROCHE DIABETES CARE AG
- Product Code
- LZG
- PMA / PMN Number
- NA
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
THE INCIDENT OCCURRED OUTSIDE THE UNITED STATES. INFORMATION CONTAINED WITHIN THIS REPORT IS ALL THAT IS AVAILABLE AT THIS TIME. IF FURTHER INFORMATION IS OBTAINED, IT WILL BE PROVIDED IN THE SUPPLEMENTAL REPORT.
THE SOFT COMPONENTS OF THE HOUSING ARE WORN DOWN HEAVILY. THEREFORE, MOISTURE MAY ENTER THE INSULIN PUMP AND DESTROY THE PUMP ELECTRONICS. THE BUTTONS FUNCTION WERE TESTED SUCCESSFUL AND MEET THE SPECIFICATION. THE PROBLEM MENTIONED BY THE CUSTOMER IS RELATED TO CARELESS HANDLING OF THE PRODUCT.
ON (B)(6) 2013, IT WAS REPORTED THAT THE UP AND DOWN BUTTONS ON THE PATIENT'S INFUSION DEVICE WERE DAMAGED AND NOT FUNCTIONING. NO PHYSIOLOGICAL EFFECTS WERE REPORTED. THE PATIENT DID NOT REQUIRE MEDICAL ASSISTANCE FROM A HEALTHCARE PROFESSIONAL OR SECOND PARTY TO ADDRESS THE ISSUE. INFUSION DEVICE WAS REPLACED AND WAS REQUESTED TO BE RETURNED FOR PRODUCT EVALUATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 116850 | ACCU-CHEK SPIRIT | INSULIN INFUSION DEVICE | LZG | ROCHE DIABETES CARE AG | 00700006862 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |