OT ULTRA2 METER
Report
- Report Number
- 3008382007-2013-05664
- Event Type
- Injury
- Date Received
- March 20, 2013
- Report Date
- February 26, 2013
- Manufacturer
- LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
- Product Code
- NBW
- PMA / PMN Number
- K053529
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- PATIENT
Narratives
THE METER INVOLVED WITH THIS COMPLAINT HAS BEEN RETURNED AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: THE METER HAS PASSED TESTING WITH NO FAULTS FOUND. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED.
ON (B)(6) 2013, THE LAY USER/ PATIENT CONTACTED LIFESCAN (LFS) ALLEGING HIS ONETOUCH ULTRA2 METER READ INACCURATELY HIGH COMPARED TO HIS FEELINGS/ NORMAL BLOOD GLUCOSE RESULTS. THE COMPLAINT WAS CLASSIFIED BASED ON THE CUSTOMER CARE ADVOCATE (CCA) DOCUMENTATION. THE ALLEGED ISSUE BEGAN A COUPLE WEEKS PRIOR TO CONTACTING LFS. THE PATIENT REPORTED BLOOD GLUCOSE RESULTS OF "148, 135, 136, 150, 171, 164 AND 80S MG/DL" WITH THE SUBJECT METER. THE PATIENT MANAGES HIS DIABETES WITH HUMALOG INSULIN AND LANTUS INSULIN. THE PATIENT REPORTEDLY CONTINUED TO ADMINISTER HIS USUAL DOSE OF MEDICATIONS. AFTER THE START OF THE ALLEGED ISSUE, THE PATIENT CLAIMS HE FELT SYMPTOMS OF LIGHT HEADED AND SWEATING. THE PATIENT DENIED RECEIVING MEDICAL TREATMENT. AT THE TIME OF TROUBLESHOOTING, THE CCA NOTED THE SUBJECT METER WAS SET TO THE CORRECT UNIT OF MEASUREMENT. QUALITY CONTROL TESTING WAS PERFORMED WHICH TESTED WITHIN SPECIFICATIONS. REPLACEMENT PRODUCTS WERE SENT TO THE PATIENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE PATIENT REPORTEDLY DEVELOPED SYMPTOMS SUGGESTIVE OF A SERIOUS INJURY AFTER THE ALLEGED METER ISSUE BEGAN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 116882 | OT ULTRA2 METER | GLUCOSE MONITORING SYS/KIT | NBW | LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL | 3345900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 77 YR | Life Threatening |