FDA Adverse Event
Malfunction
Summary report: N
OT VERIO METER
MDR report key: 3011982
·
Received March 20, 2013
Report
- Report Number
- 3008382007-2013-05654
- Event Type
- Malfunction
- Date Received
- March 20, 2013
- Report Date
- February 21, 2013
- Manufacturer
- LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
- Product Code
- NBW
- PMA / PMN Number
- K093745
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
FOLLOW-UP # 1 ¿ (07/24/2013) THE PATIENT¿S METER HAS BEEN RETURNED AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS ON (B)(4) 2013 AND (B)(4) 2013, RESPECTIVELY, WITH THE FOLLOWING FINDINGS: THE METER INVOLVED WITH THIS COMPLAINT FAILED TESTING. THE METER WAS FOUND TO HAVE A BROKEN LCD DISPLAY. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.
Description of Event or Problem · 1
ON (B)(6) 2013, THE REPORTER CONTACTED LIFESCAN (B)(4), ALLEGING CRACKED DISPLAY. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 116505 | OT VERIO METER | GLUCOSE MONITORING SYS/KIT | NBW | LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL | 3277164 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 56 YR |