FDA Adverse Event Malfunction Summary report: N

HOMECHOICE PRO

MDR report key: 3011946 · Received March 19, 2013

Report

Report Number
1416980-2013-06652
Event Type
Malfunction
Date Received
March 19, 2013
Date of Event
January 22, 2013
Report Date
February 26, 2013
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FKX
PMA / PMN Number
K102936
Removal / Correction Number
1423500-01/08/10-001-C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PL
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION: THE DEVICE WAS RETURNED TO BAXTER, AND THE EVALUATION IS COMPLETE. THIS COMPLAINT IS AN ANCILLARY SERVICE EVENT. THE REPORTED DIFFICULTY OF AN IIPV EVENT WAS CONFIRMED THROUGH AN EVENT HISTORY LOG REVIEW. THE CAUSE WAS A FALSE EMPTY DETECT AND USE ERROR, INAPPROPRIATE BYPASS OF THE LOW DRAIN VOLUME ALARM, NOT INCLUDING THE I-DRAIN. THE HOMECHOICE APD SYSTEMS PATIENT AT-HOME GUIDE DESCRIBES THE PROCEDURE FOR BYPASSING A LOW DRAIN VOLUME ALARM. THE TEXT ON PAGES 18-45 AND 18-46 INCLUDES, "BYPASSING A LOW DRAIN VOLUME ALARM CAN LEAVE FLUID IN THE PERITONEAL CAVITY AND RESULT IN AN INCREASED INTRAPERITONEAL VOLUME (IIPV) SITUATION." A REVIEW OF THE LABEL FOR THE PRODUCT FAMILY WILL BE CONDUCTED. IF THERE IS ANY FURTHER RELEVANT INFORMATION FROM THAT REVIEW, A SUPPLEMENTAL MEDWATCH WILL BE FILED.

Description of Event or Problem · 1

DURING THE EVALUATION OF A RETURNED HOMECHOICE MACHINE, ONE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED WHICH OCCURRED DURING THE THERAPY INITIATED ON (B)(6) 2013, AT 22:50:01. DURING NIGHT DRAIN CYCLE SIX, THE PATIENT'S ULTRAFILTRATION READING WAS 1188ML, INDICATING THE HOME PATIENT (HP) DRAINED 1188ML MORE THAN THEIR MAXIMUM PROGRAMMED FILL VOLUME OF 1600ML. THIS INFORMATION MEETS IIPV CRITERIA. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
114058 HOMECHOICE PRO K102936 FKX BAXTER HEALTHCARE - LARGO

Patients

Seq Age Sex Outcome Treatment
1