FDA Adverse Event Malfunction Summary report: N

STERIL COV F/TRS

MDR report key: 3011920 · Received March 19, 2013

Report

Report Number
8030965-2013-00986
Event Type
Malfunction
Date Received
March 19, 2013
Date of Event
January 16, 2013
Report Date
February 22, 2013
Manufacturer
SYNTHES GMBH
Product Code
HWE
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. SUBJECT DEVICE HAS BEEN RECEIVED AND IS CURRENTLY IN THE EVALUATION PROCESS. INVESTIGATION IS ON GOING; NO CONCLUSION COULD BE DRAWN. MANUFACTURING DOCUMENTS WERE REVIEWED AND NO COMPLAINT RELATED ISSUES WERE FOUND.

Additional Manufacturer Narrative · 1

DEVICE IS USED FOR TREATMENT, NOT DIAGNOSIS. THE OILY, WHITE SUBSTANCE WAS EVALUATED BY BIOSERVICE SCIENTIFIC LABORATORIES IN GERMANY WITH THE FOLLOWING RESULTS: UNDER THE CONDITIONS OF THE PRESENT STUDY IT CAN BE STATED THAT THE POLAR EXTRACT OF THE TEST ITEM CAUSED NO SIGNS OF IRRITATION. UNDER THE CONDITIONS OF THE PRESENT STUDY IT CAN BE STATED THAT THE NONPOLAR EXTRACT OF THE TEST ITEM CAUSED SIGNS OF IRRITATION. THE PRIMARY IRRITATION INDEX FOR THE NONPOLAR TEST ITEM EXTRACT COMPARED TO THE NONPOLAR REAGENT CONTROL WAS 0.2. AS THE PRIMARY IRRITATION INDEX FOR BOTH TEST ITEM EXTRACTS WAS LESS THAN 1, THE REQUIREMENTS OF THE TEST ARE MET AND THE TEST ITEM IS CLASSIFIED AS NOT IRRITANT. PLACEHOLDER.

Description of Event or Problem · 1

THIS REPORT IS 5 OF 5 FOR COMPLAINT (B)(4).

Description of Event or Problem · 1

DEVICE REPORT FROM SYNTHES (B)(6) REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2013 AN OILY WHITE RESIDUE WAS FOUND ON END OF THE TRS ATTACHMENTS AND SPLATTERED INSIDE OF THE STERILIZING PAPER, POST THE STERILIZATION PROCESS. NO FURTHER INFORMATION IS AVAILABLE AT THIS TIME. THIS IS 5 OF 5 REPORTS FOR THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
113164 STERIL COV F/TRS HWE SYNTHES GMBH 1000008

Patients

Seq Age Sex Outcome Treatment
1